期刊文献+

信号增强型电化学免疫传感器用于NT-proBNP高灵敏检测的研究

Research on the signal enhanced electrochemical immunosensor for the detection of NT-proBNP
下载PDF
导出
摘要 目的、构建用于人N末端前钠尿肽(NT-proBNP)高灵敏检测的电化学传感器。方法采用自组装方法将牛血清清蛋白-碳纳米管复合物(BSA-MWNTs)田栽在洁净的玻碳龟板表面,结合共价交联固定化技术和亲和素-生物素信号放大技术,制得NT-proBNP高灵敏免疫传感器,优化实验条件,采用电化学方法对电极的响应特性进行研究。结果在最优实验条件下,该传感器对NT-proBNP响应的线性范围为1-30ng/mL和30-150ng/mL,检出限为0.03ng/mL。结论该方法可实现对NT-proBNP的超灵敏检测。 Objective To construct an electrochemical immunosensor for detection NT-proBNP with high sensitivity. Methods The proposed immunosensor was prepared by the follows:firstly,the hybrid material BSA- MWNTs was self-assembled onto the surface of cleaned glass carbon electrode;subsequently,streptavidin was covalently linked to the modified electrode;then, biotin labeled antibody could be absorbed to the electrode by the reaction between streptavidin and biotin. The effects of experimental paramenters have been investigated. Results Under the optimal experiment conditions,the prepared immunosensor displayed a wide liner relationship between 1- 30 ng/mL and 30-150.0 ng/mL with a detection limit of 0.03 ng/mL. Conclusion The NT- proBNP imunosensor can be utilized for the detection of NT-proBNP with low concentration.
作者 苏会岚
出处 《重庆医学》 CAS 北大核心 2016年第26期3684-3687,共4页 Chongqing medicine
基金 国家自然科学基金资助项目(81401757)
关键词 电化学 免疫传感器 NT-PROBNP 牛血清清蛋白 碳纳米管 electrochemistry imhauno-sensor NT-proBNP bovine serum albumin carbon nanOtube
  • 相关文献

参考文献15

  • 1Emdin M, Passino C, Prontera C, et al. Comparison of brain natriuretic peptide(BNP) and amino-terminal ProB- NP for early diagnosis of heart failure[J]. Clin Chem, 2007,53(7) : 1289-1297.
  • 2Kim JY, Lee EY, Jee J H, et al. N-terminal pro-brain natri- uretic peptide (NT-proBNP) in Type 2 diabetes with left ventricular dysfunction[J]. Diabetes Res Clin Pract, 2007,77 Suppl 1:$238-242.
  • 3易维京.心衰标志物NT-proBNP高效免疫传感器检测的关键技术研究[Z],2011.
  • 4Liang W,Li Y,Zhang B,et al. A novel microfluidic immu- noassay system based on electrochemical immunosensors: an application for the detection of NT-proBNP in whole blood[J]. Biosens Bioelectron, 2012,31 ( 1 ) : 480-485.
  • 5毛燕群.快速灵敏检测HCV核心抗原的免疫新方法及其性能分析[J].重庆医学,2012,41(7):701-703. 被引量:4
  • 6Zhuo Y, Yi WJ, Lian WB, et al. Ultrasensitive electro- chemical strategy for NT-proBNP detection with Gold nanochains and horseradish peroxidase complex amplifi- cation[J]. Biosens Bioelectron, 2011,26 (5) : 2188-2193.
  • 7Yu S,Zou G,Wei Q. Ultrasensitive electrochemical immu- nosensor for quantitative detection of tumor specific growth factor by using Ag@ CeO2 nanocomposite as la- bels, Talanta[Z], 2016 : 11-17.
  • 8Zhang H, Ma L, Li P, et al. A novel electrochemical im- munosensor based on nonenzymatic Ag@Au-Fe304 nano- electrocatalyst for protein biomarker detection[J]. Biosens Bioelectron, 2016,85 : 343-350.
  • 9逯岭松,刘蓓,马霄,虞成,吴舜,冷建杭.电化学免疫传感器超灵敏检测髓过氧化物酶的研究[J].重庆医学,2015,44(36):5109-5111. 被引量:3
  • 10Zhang A, Xiang H, Zhang X, et al. A novel sandwich elec-trochemiluminescence immunosensor for ultrasensitive detection of carbohydrate antigen 19-9 based on immobili- zing luminol on Ag@BSA core/shell microspheres[J]. Bi- osens Bioelectron, 2016,75 : 206-212.

二级参考文献24

  • 1洪俊,饶永彩.HCV核心抗原测定用于丙型病毒性肝炎早期诊断临床价值的评估[J].临床检验杂志,2005,23(2):133-133. 被引量:17
  • 2FRENS G.Controlled nucleation for the regulation of theparticle size in monodisperse gold suspensions[J].NatPhys Sci,1973,241:20-22.
  • 3Huang KJ,Niu DJ,Xie WZ,et al.A disposable electro-chemical immunosensor for carcinoembryonic antigenbased on nano-Au/multi-walled carbon nanotubes-chi-tosans nanocomposite film modified glassy carbon elec-trode[J].Anal Chim Acta,2010,659(1/2):102-108.
  • 4Kashiwakuma T,Hasegawa A,Kajita T,et al.Detectionof hepatitis C virus specific core protein in serum of pa-tients by a sensitive fluorescence enzyme immunoassay(FEIA)[J].J Immunol Methods,1996,190(1):79-89.
  • 5Zhang S,Zheng F,Wu Z,et al.Highly sensitive electro-chemical detection of immunospecies based on combina-tion of Fc label and PPD film/gold nanoparticle amplifica-tion[J].Biosens Bioelectron,2008,24(1):129-135.
  • 6卢业成,江振友,邝燕玲,陈万山,谭奕洲,李涤蓉.游离丙型肝炎病毒核心抗原检测的临床价值探讨[J].细胞与分子免疫学杂志,2007,23(7):635-637. 被引量:10
  • 7Kataoka Y, Shao M, Wolski K, et al. Myeloperoxidase lev- els predict accelerated progression of coronary atheroscle- rosis in diabetic patients: insights from intravascular ul- trasound[J]. Atheroselerosis, 2014,232 (2) : 377-383.
  • 8Brennan ML, Penn MS, Van Lente F, et al. Prognostic value of myeloperoxidase in patients with chest pain[J]. N Engl J Med,2003,349(17) :1595-1604.
  • 9Frossard M, Fuchs I, Leitner JM, et al. Platelet function predicts myocardial damage in patients with acute myo- cardial infarction [J].Circulation, 2004, 110 ( 11 ) : 1392- 1397.
  • 10Sharma MK, Narayanan J, Upadhyay S, et al. Electrochemical immunosensor based on bismuth nanocomposite film and cad- mium ions functionalized titanium phosphates for the detec- tion of anthrax protective antigen toxin[J]. Biosens Bioelec- tron, 2015,74 : 299-304.

共引文献4