WS/T406—2012标准在SysmexCS-2000i血凝仪分析性能验证中的应用 被引量:1

Application of 406-2012 WS/T standard in performance verification of Sysmex CS-2000i blood coagulation analyzer
下载PDF
导出
摘要 目的评价SysmexCS-2000i全自动血凝仪的分析性能。方法根据中华人民共和国卫生行业标准WS/T406—2012文件及相关文献,对SysmexCS-2000i全自动血凝仪检测Prr、APTT和Fib的批内精密度、日间精密度、线性、正确度和准确度进行验证,确保分析性能满足临床需要。结果PT、APTT、Fib的批内精密度1.44%~3.58%,日间精密度2.28%-4.88%;Fib的标本浓度在0.39-7.79g/L范围内呈一次线性;正确度和准确度均符合WS/T406—2012标准要求。结论SysmexCS-2000i全自动血凝仪分析性能满足临床需要,WS/T406—2012行业标准具有较好的操作性和实用性,易于常规实验室使用和推广。 Objective To evaluate the analytical performance of Sysmex CS-2000i automatic blood coagulation analyzer. Methods PT, APTT, and Fib were detected by Sysmex CS-2000i automatic blood coagulation analyzer. The corresponding within-run precision, inter-day precision, linearity, trueness, and accuracy were analyzed. The WS/T 406-2012 document-a health industry standard of the People's Republic of China--was used to evaluated the aforementioned analytical performances. Results For PT, APTT, and Fib, the within- run precision and the inter-day precision were 1.44%-3.58% and 2.28%-4.88%, respectively. Fib had a good linearity within 0.32-7.79 g/L. Both the trueness and the accuracy could meet the requirement of WS/T 406-2012 standard. Conclusions The analytical performance of Sysmex CS-2000i automatic blood coagulation analyzer is able to meet the clinical requirement. The WS/T 406-2012 standard has good operability and practicability, and it is suitable to be applied and popularized in routine laboratory.
出处 《国际医药卫生导报》 2016年第18期2854-2856,2859,共4页 International Medicine and Health Guidance News
关键词 全自动血凝仪 分析性能 WS/T 406—2012 Automatic blood coagulation analyzer Performance analysis WS/T 406-2012
  • 相关文献

参考文献7

二级参考文献26

共引文献93

同被引文献21

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部