摘要
目的:研究多层螺旋CT儿童鼻窦低剂量扫描的临床应用。方法:选取80例需鼻窦扫描患儿,随机分成低剂量组和常规组,分别进行低剂量和常规剂量CT扫描,评价图像质量。结果:低剂量组DLP值明显低于常规组(P<0.05),低剂量组图像质量评分略低于常规组(P>0.05)。结论:低剂量扫描能够在不影响临床诊断的同时,降低患儿辐射剂量。
Objective:The clinical application of multi-slice spiral CT in nasal sinuses of children with low dose was studied. Methods: 80 patients were randomly divided into low dose group and routine group respectively. The image quality was evaluated by low dose and conventional dose of CT scanning. Results:Low dose group DLP was signiifcantly lower than the conventional group (P〈0.05), image quality of low dose group score was slightly lower than the conventional group (P〉0.05). Conclusion:Low dose scanning didn’t affect the clinical diagnosis, moreover, reduce the radiation dose.
出处
《中国医疗器械信息》
2016年第9期50-51,80,共3页
China Medical Device Information
基金
齐齐哈尔市科学技术计划项目
项目编号:SFGG201526
关键词
低剂量
儿童
鼻窦
Low dose
Children
Nasal Sinus