SYSMEX XE-5000全自动血液分析仪在儿童脑脊液常规检测中的应用 被引量:2

Application of SYSMEX XE-5000 automatic blood analyzer for the routine determination of cerebrospinal fluid among children
下载PDF
导出
摘要 目的探讨SYSMEX XE-5000全自动血液分析仪(简称XE-5000)在儿童脑脊液常规检测中的应用价值。方法选取194份住院患儿的脑脊液标本,采用XE-5000体液模式进行脑脊液常规检测,同时行人工显微镜镜检,比较仪器法与人工镜检法检测结果的符合性。结果当白细胞(WBC)计数≥30个/μL时,2种方法差异无统计学意义(P>0.05)。当WBC计数<30个/μL时,2种方法差异有统计学意义(P<0.05)。当WBC计数≥10个/μL时,2种方法之间单个核细胞计数及多个核细胞计数均具有良好的相关性(P<0.01)。当红细胞(RBC)计数>1 000个/μL时,2种方法具有良好的相关性(r=0.98,P<0.01)。结论应用XE-5000体液模式检测儿童脑脊液,在WBC总数≥30个/μL时具有较好的应用价值,可代替人工镜检。 ObjectiveTo investigate the signiflcance of SYSMEX XE-5000 automatic blood analyzer for the routine determination of cerebrospinal fluid among children. MethodsA total of 194 cerebrospinal fluid samples were collected and determined by XE-5000 and manual microscopy. The consistency was evaluated. ResultsWhen white blood cell(WBC) count was ≥30 cells/μL,there was no statistical signiflcance between the 2 methods(P〉0.05). However,when WBC count was 〈30 cells/μL,there was a statistical signiflcance between the 2 methods(P〈0.05). When WBC count was ≥10 cells/μL,there were good correlations in counting mononuclear cells and multinuclear cells between the 2 methods (P〈0.01). When red blood cell(RBC) count was 〉1 000 cells/μL,there was a good correlation between the 2 methods(r=0.98,P〈0.01). ConclusionsXE-5000 can be used in the determination of cerebrospinal fluid among children,when WBC count is ≥30 cells/μL.
出处 《检验医学》 CAS 2016年第9期807-809,共3页 Laboratory Medicine
关键词 细胞计数 脑脊液 儿童 SYSMEX XE-5000全自动血液分析仪 Cell counting Cerebrospinal fluid Children SYSMEX XE-5000 automatic blood analyzer
  • 相关文献

参考文献7

二级参考文献25

  • 1李丽,王明元,程根林,徐惠新,周群刚,夏毓伟,曹维娟,张辉.流式细胞术在成分血残留白细胞检测中的应用[J].临床输血与检验,2004,6(3):177-179. 被引量:6
  • 2丛玉隆,秦小玲.既要发展现代技术 也要继承经典方法[J].中华检验医学杂志,2005,28(2):129-130. 被引量:53
  • 3束国防,高茂馗,芦慧霞.XE-2100血液分析仪的应用评价[J].现代医学,2005,33(5):323-325. 被引量:5
  • 4杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127.
  • 5Aulesa C, Pastor I, Naranjo D, et al.Validation of the CoulterLH750 in a hospital reference laboratory[J]. Lab Hematol,2003,9: 15-28.
  • 6Ziebig R, Lun A, Sinha P. Leukocyte counts in cerebrospinal fluid with the automated hematology analyzer CellDyn 3500 and the urine flow cytometer UF-100 [J]. Clin Chem, 2000, 46(2) :242-247.
  • 7Aune MW, Sandberg S. Automated counting of white and red blood cells in the cerebrospinal fluid[ J]. Clin Lab Hacmatol, 2000,22 (4) : 203-210.
  • 8Van Acker JT, Delanghe JR, Langlois MR, et al. Automated flow cytometric analysis of cerebrospinal fluid [J]. Clin Chem, 2001,47(3) :556-560.
  • 9Aune MW, Becker JL, Bmgnara C, et al. Automated flow cytometric analysis of blood cells in cerebrospinal fluid: analytic performance [J]. Am J Clin Pathol, 2004,121 ( 5 ) :690-700.
  • 10Hoffmann JJ, Janssen WC. Automated ciunting of cells in cerebrospinal fluid using the CellDyn-4000 haematology analyzer [ J ]. Clin Chem Lab Med, 2002,40 (11) :1168-1173.

共引文献65

同被引文献17

引证文献2

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部