期刊文献+

科氏质量流量计振子平衡特性对零点漂移的影响研究 被引量:2

Research on the Influence of Balance Characteristics of Oscillator of CMF on Zero Drift
下载PDF
导出
摘要 科氏流量计的零点漂移问题是制约产品测量精度和性能稳定性的主要瓶颈。文章研究了振子的平衡特性(包括:激振器位置不对称、检测器位置不对称、定距板位置不对称、测量管附加质量不对称和测量管体积不对称等)对零点的影响,对导致科氏流量计振子不平衡的因素进行了仿真分析和实验分析,获得了随着振子不平衡的加剧零点的变化趋势,对科氏流量计的结构设计和性能参数改善提供理论指导。 The problem of zero drift of Coriolis Mass Flowmeter(CMF)is one of the main constraints to the productsin the measurement of accuracy and performance stability. This paper studied the influence of the balance character-istics of oscillator to zero,including asymmetry of exciter position,asymmetry of detector position,asymmetry offixed-pitch plate position,asymmetry of additional mass in the measuring tube and asymmetry of volume of measur-ing tube. Besides,this paper also conducted a few simulation analyses and experimental analyses to the factors re-sulting in unbalanced of the oscillator of CMF,obtaining the trend of zero value with the intensification of imbalanc-es of oscillator,Analyses in this paper can provide theoretical guidance for the design and the performance parame-ters to improve of CMF.
出处 《传感技术学报》 CAS CSCD 北大核心 2016年第9期1311-1316,共6页 Chinese Journal of Sensors and Actuators
基金 国家自然科学基金项目(60974109)
关键词 科氏质量流量计 零点漂移 ANSYS WORKBENCH 不平衡 Coriolis mass flowmeter zero drift ANSYS Workbench unbalance
  • 相关文献

参考文献11

二级参考文献71

共引文献101

同被引文献20

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部