摘要
目的采用超高效液相色谱法测定人血清中伏立康唑(VRC)浓度,并用于临床上该药的血药浓度监测。方法以地西泮为内标,色谱柱为Agilent Poroshell 120 EC-C18(4.6×100 mm,2.7μm),流速1 ml/min,流动相为乙腈-0.2%甲酸水溶液(45∶55),检测波长254 nm,进样量10μl,柱温40℃。结果 VRC及地西泮的保留时间分别为2.52和4.06 min;血清中VRC线性范围为0.12-12.0μg/ml(r=0.9999),定量下限为0.12μg/ml,日内精密度相对标准偏差(RSD)均〈1.9%,日间精密度RSD均〈2.1%,低、中、高3个浓度方法回收率为98.39%、101.16%和96.63%。结论该方法用于VRC的血药浓度检测,具有操作简便、稳定性良好,节约时间等优点。
Objective To determinate the concentration of Voriconazole in human serum by ultra performance liquid chromatography(UPLC). Methods Chromatography was performed on Agilent Poroshell 120EC-C18(4.6 × 100 mm, 2.7 μm), the internal standard was Diazepam. The flow rate was 1 ml/min, the mobile phase was acetonitrile-water(0.2% formic acid)(45 : 55), and the detection wavelength was 254 nm. Injection volume was 10 μl and the column temperature was 40℃. Results The retention time of Voriconazole and the internal standard Diazepam was 2.52 min and 4.06 min, respectively. The calibration curve was linear over the concentration range of 0.12-12.0 μg/ml(r = 0.9999) with the lower limit of quantitation(LLOQ) at 0.12 μg/ml.Intra- and inter-day RSD was below 1.9% and 2.1%, respectively. The extraction recovery rate of the low,medium and high concentrations was 98.39%, 101.16% and 96.63%, respectively. Conclusions The UPLC is a simple, highly selective and sensitive method for determining serum level of Voriconazole.
出处
《中国现代医学杂志》
CAS
北大核心
2016年第19期37-41,共5页
China Journal of Modern Medicine
基金
安徽医科大学第二附属医院博士科研启动基金(No:2012BKJ022)
安徽医科大学科学研究基金(No:2015xkj020)
关键词
伏立康唑
超高效液相色谱法
血药浓度监测
Voriconazole
ultra performance liquid chromatography
therapeutic drug monitoring