摘要
针对九江二井大流量的特点,开展井口装置改造,数字化氡观测数据得到显著提升。本文通过设计实验,采用新型恒流装置、脱气装置与P2000测氡仪组成的观测系统定量研究水流量对数字化氡观测数据的影响。结果表明,在一定范围内氡观测值与水流量呈正比关系。
According to the characteristics of large flow in Jiujiang No.2 well, well-head equipment was reformed and digital observation of radon got promoted significantly.This paper discusses quantitive relationship between water flow with digital observation of radon through experiment, by the new constant current device, degassing device and P2000 measuring radon instrument.The results show that the numerical value of radon is proportional to water flow in a certain range.
出处
《防灾减灾学报》
2016年第3期30-33,共4页
Journal of Disaster Prevention And Reduction
基金
中国地震局监测预报司流体氡观测仪器对比实验研究项目
关键词
流量
氡
观测系统
water flow
radon
observation system