期刊文献+

超快速液相色谱-串联质谱法检测人血清中雌酮、雌二醇及雌三醇的方法学研究 被引量:8

Study on the determination of estrone,estradiol and estriol in human serum by UFLC-MS/MS method with pre-column derivatization
下载PDF
导出
摘要 建立灵敏、可靠、稳定的柱前衍生化超快速液相色谱-串联质谱(ultra-fast liquid chromatography-mass spectrometry,UFLC-MS/MS)方法检测人血清中超微量雌酮、雌二醇和雌三醇。方法:取200μL血清样品,采用两步液萃取法,通过丹磺酰氯衍生化后进行样品分析,以雌二醇-d3作为内标。采用Welch Ultimate UHPLC XB-C18色谱柱(2.1mm×50mm,1.8μm),乙腈-水(含0.1%甲酸)为流动相,梯度洗脱,流速0.20mL/min,柱温40℃,进样量10μL。在电喷雾离子(electrospray ionization,ESI)源正离子模式下,选择多反应监测(multiple reaction monitoring,MRM)模式。经检测,雌酮、雌二醇、雌三醇及内标的检测离子对分别为:m/z为504.0→171.0、m/z为506.0→171.0、m/z为522.0→171.0和m/z为509.0→171.0。结果:丹磺酰氯衍生法可大幅度提高雌酮、雌二醇、雌三醇在ESI源下的离子化效率,血清中3种成分的线性范围均为0.01-5.00μg/L,定量下限均达到10ng/L,样品回收率50%以上,批内、批间准确度与精密度均小于15%,满足生物样品的分析要求。结论:该方法灵敏度高、重现性好,可以用于临床血清样品中雌激素水平的定量分析。 A selective and sensitive ultra-fast liquid chromatography-mass spectrometry (UFLC-MS/MS) method with pre-column derivatization was developed for the simultaneous determination of estrone, estradiol and estriol in human serum. The analytes were extracted from 200 μL serum samples by liquid-liquid extraction using two-step extraction method via methyl tert-butyl ether and dichloromethane. And then derivatization process was carried out using dansyl chloride. Estradiol-d3 was selected as the internal standard. Chromatographic separation was achieved on a Welch Ultimate UHPLCXB-C18 column (2. 1 mm× 50 mm, 1.8 μm) using mobile phase consisted of aeetonitrile and water containing 0. 1% formic acid. The column temperature was 40℃. The flow rate was set at 0. 20 mL/min and the injection volume was 10μL. Detection and quantification were performed using a mass spectrometer by the multiple reaction monitoring (MRM) in positive electrospray ionization mode. The transition monitored were m/z 504. 0→171.0 for estrone, m/z 506. 0→171.0 for estradiol, m/z 522. 0→171.0 for estriol and m/z 509.0→171.0 for IS, respectively. The calibration curves of all three analytes in serum show good linearity over the concentration range of 0. 01-5.00 μg/L with a limit of quantitation at 10 ng/L. The extraction recoveries are 〉50% as well as both the intra-and intevday accuracy and precision are within 15%. The sensitive and reproducible method could be successfully applied for the quantification of estrone, estradiol and estriol in clinical serum samples.
出处 《中国科技论文》 CAS 北大核心 2016年第18期2062-2068,共7页 China Sciencepaper
基金 国家自然科学基金资助项目(81302835) 高等学校博士学科点专项科研基金资助项目(20133221120010)
关键词 药物分析 LC-MS/MS 雌酮 雌二醇 雌三醇 柱前衍生化 pharmaceutical analysis LC-MS/MS estrone estradiol estriol pre-column derivatization
  • 相关文献

同被引文献54

引证文献8

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部