64排螺旋CT冠状动脉成像中降低对比剂用量的初步研究 被引量:1

Initial Experience in Reducing the Amount of Contrast Media in Sixty-Four-Slice Spiral CT Coronary Angiography
下载PDF
导出
摘要 目的:研究多排螺旋CT冠状动脉成像时降低对比剂剂量、注射速率及对比剂浓度的可行性。材料与方法:100例可疑或确诊冠心病而行CT冠状动脉造影的患者,随机分为五组,每组20例。对比剂剂量、注射速率、对比剂浓度分别设为A组70m L、5.0m L/s,以5ml/s速率追加30ml盐水,B组50m L、5.0m L/s,以5ml/s速率追加30ml盐水,C组50m L、3.5m L/s、以3.5ml/s速率追加30ml盐水。D组60m L、4m L/s、以4ml/s速率追加30ml盐水。E组50m L、4.5m L/s、以4.5ml/s速率追加30ml盐水。A、B组,扫描触发点设定于主动脉弓;C、D、E组,扫描触发点设定于降主动脉。应用美国心脏协会(AHA)冠状动脉改良分段法进行冠状动脉成像质量评价。测量各段冠状动脉管腔内CT值。分别比较各方法获得的冠状动脉成像质量与管腔内CT值。使用SPSS14.0行统计学分析,采用秩和检验Wilcoxon法,当P<0.05时,差异有统计学意义。结果:图像成像质量比较,A组与D组比较,P>0.05,两组间差异没有统计学意义。B、C、E各组间的成像质量比较,P>0.05,差异没有统计学意义。A、D与其余三组间差异有统计学意义,P<0.05。结论:注射速率4ml/s、总量60ml追加生理盐水30ml的对比剂注射方案,能够在减少对比剂用量的同时,保持与应用70ml对比剂相同的诊断效能。 Purpose: To evaluate how to decrease the Iodine intake of patients by using the least contrast quantity and iodine concentration in 64-detector computed tomography coronary imaging. Materials and Methods: 100 patients underwent CT coronary angiography using 64-detector CT who were known or suspected as CAD was randomly divided into five groups: A, B, C, D and E. Contrast quantity, velocity of injection and iodine concentration was respectively used as 70ml contrast (5.0mL/s) and 30ml saline(5.0ml/s) in group A; 5Oral contrast (5.0mL/s) and 30ml saline(5ml/s) in group B; 50ml contrast (3.5mL/s) and 3Oral saline(3.5ml/s) in group C; 60ml contrast (4. 0mL/s) and 30ml saline(4ml/s) in group D; 50ml contrast (4. 5mL/s) and 3Oral saline(4. 5ml/s) in group E. Image qualities of all coronary images according to the AHA classification were evaluated respectively. CT values of the cavity were measured respectively. Image qualities and CT values of every group were compared by using Wilcoxon rank sum test respectively. There was significant difference when P 〈 0.05. Results: There was no significant difference between the image quality of group A and D. There was no significant difference between the image quality of group B, C and E. There was no significant difference between the CT values of 50ml contrast groups. There was significant difference between the CT values of AD and BCE groups. Conclusion: Images with the same quality could be obtained by using lower contrast amount, velocity of injection and iodine concentration, such as 60ml contrast (4. 0mL/s) and 30ml saline(4ml/s).
出处 《现代医用影像学》 2016年第5期872-874,共3页 Modern Medical Imageology
关键词 冠状动脉 对比剂 体层摄影术 X线计算机 coronary artery, contrast media, tomography, X-ray computed
  • 相关文献

参考文献2

二级参考文献13

  • 1朱海峰,范家栋,郑卓肇,何汇忱,舒邦明.16层螺旋CT冠状动脉成像初步探讨[J].中华放射学杂志,2006,40(10):1064-1068. 被引量:24
  • 2[1]Hong C., Backer C., Huber A., et al. ECG-gated reconstructed multi-detector row CT Coronary angiography: Effect of varying trigger delay on image quality. Radiology, 2001,220:712
  • 3[3]GerberT. C., KuzoR.S., LaneG. E., et al. Image quality in a standartized algorithm for minimally invasive coronary angiography with multislice spiral computed tomography. Journal of Computer Assisted Tomography, 2003, 27(1): 62
  • 4[4]Achenbach S., Giesler T., Ropers D., et al. Detection of coronary artery stenoses by contrast-enhanced retrospectively electro-cardiographically-gated,multislice spiral computed tomography. Circulation, 2001,103:2535
  • 5[5]Achenbach S., Ulzheimer S., Baumu U., et al. Noinvasive coronary angiography by retrospectively ECG-Gated multislice spiral CT. Circulation, 2000,102:2823
  • 6[6]Kopp A.F., Schroeder S., Kuettner A., et al Coronary arteries: Retro spectively ECG-Gated multi-detector row CT angiography with selective opti mization of the image reconstruction window. Radiology, 2001,221:683
  • 7Heuschmid M, Kuettner A, Schroeder S , et al. ECG-gated 16-MDCT of the coronary arteries : assessment of image quality and accuracy in detecting stenoses. AJR ,2005, 184:1413-1419.
  • 8Suzuki S, Furui S, Kaminaga T. Accuracy of automated CT angiography measurement of vascular diameter in phantoms: effect of size of display field of view, density of contrast medium, and wall thickness. AJR, 2065,184 : 1940-1944.
  • 9国家药典委员会编.中华人民共和国药典.临床用药须知,化学药和生物制品卷.北京:人民卫生出版社,2005.
  • 10Sudhir K, MacGregor JS, Barbant SD, et al. Assessment of coronary conductance and resistance vessel reactivity in response to nitroglycerin, ergonovine and adenosine: in vivo studies with simultaneous intravascular two-dimensional and Dopperultrasound. J Am Coll Cardiol, 1993,21 : 1261-1268.

共引文献60

同被引文献2

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部