采用线性质控品评价Coulter LH780血液分析仪的线性

Evaluation of the linearity of Coulter hematology analyzer LH 780
下载PDF
导出
摘要 目的:对Coulter LH 780全自动血液分析仪做线性评价。方法:采用原装配套试剂、校准品和室内质控品,以原装线性质控品为样本评价白细胞计数(WBC)、红细胞计数(RBC)、血红蛋白浓度(Hgb)以及血小板浓度(Plt)4项参数的线性范围。结果:4项参数的检测结果均呈良好线性关系,WBC、RBC、Hgb及Plt的线性范围分别为(0.03~363.5)×10~9/L、(0.00~7.92)×10^(12)/L、0.3~236.0 g/L以及(0.3~2511.0)×10~9/L。结论:LH 780血液分析仪的线性范围宽泛,能满足临床实验室的检测要求。 Objective:To evaluate the linearity of Coulter automated hematology analyzer LH 780. Methods:The linearity control used to be samples, the linear range of WBC, RBC, Hgb, Plt were evaluated with Coulter original reagents, calibrator and control. Results:The test results of the four parameters showed good linear relationship. The linear range of WBC, RBC, Hgb, Plt were (0.03~363.5)×109 /L, (0.00~7.92)×1012/L, 0.3~236.0 g/L and (0.3~2511.0)×109/L, respectively. Conclusion:For Coulter hematology analyzer LH 780, there is a wide linearity that can meet the demands of clinical laboratories.
出处 《中国医学装备》 2016年第7期22-25,共4页 China Medical Equipment
基金 国家"十二五"重大新药创制科技专项(2012ZX09303-008-001)"心血管创新药物临床研究技术平台建设"
关键词 血液分析仪 线性 白细胞 红细胞 血红蛋白 血小板 Hematology analyzer Linearity White blood cell count Red blood cell count Hemoglobin concentration Platelet concentration