摘要
我国原有铀矿γ测井模型标准铀含量上限为1.09%,应用其校准的γ测井仪对高品位铀矿的测量结果曾饱受争议。在2014年研建高品位(1.83%、5.09%)铀模型后,用其对γ测井仪的计量特性在进行了试验验证后,结果表明仪器死时间和铀含量灵敏度没有显著变化,说明γ测井仪在铀含量5%以内的测量结果仍然可靠。
The upper range of uranium concentration of our countrys' γ logging pad is 1.09% ,the measurement result of high uranium concentration deposit by γ logger that calibrated by 3, logging pads was controversial. After researched and built the calibration pads of high concentration of uranium ( 1.83% ,5.09% ) in 2014, carried out the test of metrologieal characteristics of γ logger through the pads, proved there is not notable change of dead time and sensitivity of uranium content of γ logger, indi- cated γ loggers' measuring result is reliable when uranium concentration under 5%.
出处
《宇航计测技术》
CSCD
2016年第5期54-57,共4页
Journal of Astronautic Metrology and Measurement
基金
核工业地质局铀矿地质勘查科研项目资助
项目编号:201582
关键词
高品位铀模型
γ测井仪
校准
High uranium concentration of calibration pad γ logger Calibration