准绝热法微型镓固定点相变特性研究 被引量:8

Research on Phase Change Characteristic of Miniature Gallium Fixed-point Using Quasi-adiabatic Measurement Method
下载PDF
导出
摘要 介绍了可用于原位校准的微型镓相变固定点形成过程,准绝热测试系统的结构和工作原理,测量了微型镓相变固定点在准绝热测试系统中的相变特性。实验结果表明微型相变镓固定点的相变温度与加热功率之间存在强线性关系,其线性拟合因子R^2为0.9957,并且随着功率越降低,其相变熔化温度越接近镓固定点的标准值。在相同的准绝热复现条件下,其相变温度的复现性优于2 mK。 The formation process of miniature Ga fixed-point which can be used for in situ calibration, the structure and working principle of the quasi-adiabatic testing system are described. The phase change characteristic of miniature Ga fixed-point is measured using this quasi-adiabatic testing system. The results show that the melting point of Ga fixed-point has strong linear relationship with the heating power. The linear fitting factor R2 is 0. 995 7. The lower the heating poweris, the closer to the standard temperature of Ga fixed-point it becomes. Further more, the repeatability of the melting points of Ga is better than 2 mK under the same quasi-adiabatic testing system.
出处 《计量学报》 CSCD 北大核心 2017年第1期19-22,共4页 Acta Metrologica Sinica
基金 国家自然科学基金(11475162) 国家863计划(2015AA123701)
关键词 计量学 镓固定点 相变特性 准绝热系统 原位校准 metrology gallium fixed-point phase change characteristic quasi-adiabatic system in situ calibration
  • 相关文献

参考文献4

二级参考文献34

  • 1卢小丰,原遵东,柏成玉,王景辉,王鹏程.基准光电高温计的光谱响应度测量[J].计量学报,2007,28(z1):65-68. 被引量:6
  • 2黄东涛,陆家钦,段宇宁,原遵东,赵琪.黑体空腔发射率计算的蒙特卡罗模型[J].清华大学学报(自然科学版),1997,37(2):28-31. 被引量:17
  • 3Gutschwager B, Schiller S, Hartmann J, et al. A lens- free InGaAs transfer radiation thermometer with an extended calibration range above 960°C [ C ]// Proceedings TEMPMEKO'04. Croatia,2004.
  • 4Sakuma F, Ma L, Suzuki T, et al. Calibration and characterization of Chino 900 nm silicon narrow - band radiation thermometer [ C ]//SICE Annual Conference in Sapporo. Sapporo, 2004, 72 - 76.
  • 5Preston-Thomas H. The International Temperature Scale of 1990( ITS--90 ) [ J ]. Metrologia, 1990, 27 ( 2 ) : 107 - 108.
  • 6Ancsin J, Phillips J. Argon Triple Point Realization Cryostat for Platinum Resistance Long Stem Thermometers [ J ]. Review of Scientifw Instruments, 1976,47 ( 12 ) : 1519 - 1521.
  • 7Bloembergen P, Bonnier G, Ronsin H. An International Intercomparison of Argon Triple Calibration Facilities, Accommodating Long-stem Thermometers [ J ] Metrologia, 1990, 27(2) : 101 - 106.
  • 8Marcarino P, Steur P P M, Dematteis R. Realization at IMGC of the ITS--90 Fixed Points from the Argon Triple Point Upwards [ J ]. AlP Conference Proceedings,2003,684 (1) :65 -70.
  • 9Pavese F, Ancsin J, Aslrov D N, et al. An International Intercomparison of Fixed Points by Means of Sealed Cells in the Range 13.81K - 90. 686K[ J]. Metrologia, 1984, 20(4) :127 - 144.
  • 10Ahmed M G, Hermier Y, Moussa M R, et al. Argon Triple-Point Device to Calibrate Long-stem Thermometers in Quasi-adiabatic Conditions [ J ]. AIP Conference Proceedings ,2003,684( 1 ) : 197 - 202.

共引文献32

同被引文献39

引证文献8

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部