摘要
目的:应用可视化角膜生物力学分析仪(Corvis ST)及眼反应分析仪(ORA)评估飞秒激光制瓣准分子激光原位角膜磨镶术(laser in situ keratomileusis,LASIK)术后角膜生物力学变化,并分析与其他参数的相关性。方法:收集63例飞秒制瓣LASIK手术患者,所有收集对象采用右眼数据进行分析。角膜生物力学测量仪器应用可视化角膜生物力学分析仪Corvis ST及眼反应分析仪ORA,手术前及手术后1mo进行生物力学数据采集分析。配对t检验或Mann-Whitney U检验进行手术前后生物力学对比分析。Pearson或Spearman统计学方法进行相关性分析。结果:与FS-LASIK手术前相比,手术后1stA-time,Vin,2ndA length,Vout以及Radius下降,有统计学差异(P=0.00,P=0.00,P=0.00,P=0.00,P=0.00)。2ndA-time,DA以及PD手术后增加,有统计学差异(P=0.00,P=0.00,P=0.00)。1stA length以及HC time手术后改变不明显,无统计学差异(P=0.96,P=0.08)。与FS-LASIK手术前相比,手术后CH,CRF下降,有统计学差异(P=0.00,P=0.00)。手术后1stA-time,2ndA-time,DA and Radius变化量与手术前角膜中央厚度有明显相关性(P=0.01,P=0.04,P=0.03,P=0.01)。结论:飞秒激光LASIK手术后角膜生物力学参数有较明显改变,可通过Corvis ST及ORA生物力学参数计算得出相应变化,手术后生物力学参数变化与角膜厚度相关性明显。
AIM:To investigate the changes of corneal biomechanical properties before and after femtosecond laser assisted LASIK(FS- LASIK) using Corneal Visualisation Scheimpflug Technology(Corvis ST) and Ocular Response Analyzer(ORA),and the correlation with other myopic parameters.METHODS:Sixty three patients(63 eyes) who had myopic femtosecond laser assisted LASIK(FS-LASIK)were enrolled in the study.The right eye of each patient was analyzed in this study.The corneal biomechanical parameters pre-operative and 1mo post-operative was measured with the Corvis ST(Oculus,Wetzlar,Germany)and ORA(Reichert,Buffalo,New York,USA).Comparison of the biomechanical property values before and after surgery was peformed using Paired t-test or MannWhitney U.Pearson or Spearman correlations were used to evaluate the relationship between parameters.RESULTS:The postoperative 1^(st) A-time,Vin,2^(nd) A length,Vout,HC time and Radius demonstrate significant decreases comparing with preoperative values(P=0.00,P=0.00,P=0.00,P=0.00,P=0.00,P=0.00 respectively).The postoperative 2^(nd) A-time,DA and PD significantly increases(P=0.00,P=0.00,P=0.00),however,the 1^(st) A length had no significant difference after surgery.The CH and CRF were significantly lower after FS-LASIK(P =0.00,P = 0.00).A statistically significant correlation coefficient was found between preoperative central corneal thickness(CCT) with postoperative-preoperative changes of 1^(st) A-time,2(nd) A-time,DA and Radius respectively(P=0.01,P=0.04,P=0.03,P=0.01).CONCLUSION:There were significantly changes of corneal biomechanical properties after FS-LASIK surgery.The changes of corneal biomechanical properties after FS-LASIK can be reflected by some parameters of Corvis ST and ORA.The mainly influence of corneal biomechanical alteration was possibly correlation with corneal thickness.
出处
《国际眼科杂志》
CAS
2017年第2期195-199,共5页
International Eye Science
基金
Supported by Xi'an Social Development Research Project Grant[No.SF1513(2)]
Shaanxi Social Development Research Project Grant(No.2016SF-274)
Shaanxi Provincial Health Research Project(No.2016E002)~~