图像引导放射治疗在653例宫颈癌摆位误差的分析 被引量:25

Analysis of the setup error in 653 cervix cancer patients treated with image-guided radiotherapy
下载PDF
导出
摘要 目的:应用锥形束CT(CBCT)评价宫颈癌患者图像引导放射治疗(IGRT)的摆位误差及体重因素对摆位误差的影响。方法:回顾性分析应用Trilogy直线加速器治疗的653例宫颈癌患者资料,使用其机载影像系统(OBI)行CBCT,共采集1175幅容积图像与计划CT图像进行配准,记录升降(X轴)、进出(Y轴)和左右(Z轴)3个轴向的误差。以正常体重指数(BMI=23)为分界,将患者分为正常组(BMI<23)和肥胖组(BMI≥23),对比分析两组BMI的差异。结果:X轴、Y轴及Z轴3个平移轴向的摆位误差分别为(-0.38±3.6)mm、(0.01±4.5)mm和(-0.13±2.8)mm,结果满足临床要求;提示体重因素对摆位误差有影响,其中肥胖组患者在X轴向的摆位误差大于正常组,差异有统计学意义(t=-2.446,P<0.05)。两组在Y轴及Z轴向的摆位误差无显著差异。结论:宫颈癌患者的整体摆位误差结果较好,能够满足临床要求;同时患者BMI对于X轴向的摆位误差具有一定影响,故应更加关注BMI较大患者的摆位情况。 Objective: The aim of this study was to evaluate setup errors by Cone-beam computed tomography(CBCT) method in cervical cancer patients with image-guided radiotherapy and discussthe impact of body mass index(BMI). Methods: A retrospective study was adopted to analyze 653 cervix cancer patients were treated with Trilogy linear accelerator from 2008 to 2015 in our department. CBCT was performed on a Varian Trilogy linear accelerator via on–board images system, and 1175 images were acquired and matched up with planning CT images, and then systematic(∑) and random(σ) errors in X(vertical)Y(longitudinal)and Z(lateral) directions were analyzed. The patients were divided into normal BMI group(BMI23)and overweight patients group(BMI≥23). The software SPSS 19.0 was used for data analyzing,and the differences between two groups were compared by paired samples T test and there was statistically significant if P0.05. Results: The errors in X、Y and Z direction were-0.38±3.6mm、0.01±4.5mm and-0.13±2.8mm, respectively, and they can meet the clinical demand;body weight factor had an impact on setup errors, and there was significant difference in vertical direction(t=-2.446, P0.05) between the two groups, however,there were no significantly difference in the other two directions. Conclusions: The overall setup errors results met the clinical demand; body weight factor can lead to larger setup errors in vertical direction, therefore more attention should be paid to the patients with larger BMI.
出处 《中国医学装备》 2017年第1期21-24,共4页 China Medical Equipment
关键词 宫颈癌 图像引导放射治疗 摆位误差 锥形束CT 体重指数 Cervix cancer Image-guided radiotherapy Setup error Cone-beam computed tomography Body mass index
  • 相关文献

参考文献6

二级参考文献66

  • 1傅万凯,林友金,刘利彬,黄丽娜,吴章桂.应用IGRT机载KV-CBCT校正胸腹部肿瘤调强适形放疗摆位误差的研究[J].实用癌症杂志,2010,25(5):511-513. 被引量:27
  • 2胡逸民.调强放射治疗的最新进展[J].中国医疗器械信息,2005,11(2):1-5. 被引量:14
  • 3潘才住,潘建基,陈传本.头颈部肿瘤精确放疗摆位误差的分析[J].实用癌症杂志,2006,21(3):326-328. 被引量:4
  • 4于龙珍,翟振宇,沈君姝.体部肿瘤精确放疗摆位误差分析[J].中国医学物理学杂志,2007,24(3):160-162. 被引量:36
  • 5Lee N,Puri DR,Blanco AI,et al.Intensity-modulated radiation therapy in head and neck cancers:an update[J].Head Neck,2007,9(4):387-400.
  • 6Eisbruch A.Intensity modulated radiotherapy of head-and-neck cancer:encouraging early results[J].Int J Radiat Oncol Biol Phys,2002,53(1):1-3.
  • 7Bortfeld T,Van-Herk M,Jiang SB,et al.When should systematic patient position errors in radiotherapy be corrected?[J].Phys Med Biol,2002,47:297-302.
  • 8VanHaaren PM,Bel A,Hofman P,et al.Influence of daily setup measurements and corrections on the estimated delivered dose during IMRT treatment of prostate cancer patients[J].Radiother Oncol,2009,90(3):291-298.
  • 9Hunt M.The effect of positional uncertainties on the treatment of primary nasop-harynx cancer[J].Med Phys,1989,16:456-464.
  • 10Mackie TR,Kapatoes J,Ruchala K,et al.Image guidance for percise conformal radiotherapy[J].Int J Radiat Oncol Biol Phys,2003,56(1):89-105.

共引文献231

同被引文献184

引证文献25

二级引证文献97

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部