期刊文献+

诺尔康人工耳蜗术后电诱发镫骨肌反射测试及临床应用 被引量:5

Electrically Evoked Stapedius Reflex and Clinical Trials in Post-Cochlear Implant Patients of Nurotron
下载PDF
导出
摘要 目的探讨电诱发镫骨肌反射(Electrically evoked Stapedius Reflex,ESR)的特性及其反应阈(Electrically evoked Stapedius Reflex Threshold,ESRT)与心理物理测量数值(T和C值)之间的关系,为人工耳蜗调试刺激参数的设定提供参考。方法选取植入诺尔康晨星人工耳蜗系统(CS-10A植入体)、能配合行为调试并可以准确反映其听声感受的植入者19例,采用心理物理方法测量其T值(Threshold Level)、C值(Comfortable Level)及相应电极的ESRT值,并对数据进行分析比较。结果 ESRT闽与植入者反映的C值具有良好的相关性,其关系为C=1.022×ESRT-15.143,而T值与C值、ESRT值之间均无相关性(P>0.05)。结论诺尔康晨星人工耳蜗系统调试中,可通过测量:ESRT值以帮助设定难测者的C值,并有助于动态观察植入者听神经功能。 Objective The aim of this study was to explore the characteristics of the electrically evoked stapedius reflex (ESR) and the electrically evoked stapedius reflex threshold (ESRT), and its relation with the threshold level (TL) and comfort level (CL), so that it can provide information for the parameter setting of cochlear implant (CI) mapping (programming). Methods 19 Nurotron Venus cochlear implant recipients, who could cooperate with mapping and who could well reflect the hearing perception, participated in this study. The TL and CL were obtained psychophysically. The results of TLs and CLs were statistically compared to the ESRT results.Results ESRT value had a significant correlation with CL, which was concluded as CL=l.022xESRT-15.143,but TL had no significant correlation with ESRT and CL (P〉0.05). Conclusion ESRT can be used to estimate the CL in Nurotron Venus cochlear implant users who cannot well cooperate with the mapping, and also can be used to evaluate the nerve function.
作者 银力 高珊仙 胡显亚 滕晓芳 YIN Li GAO Shan-xian ttU Xlan-ya TENG Xiao fang
出处 《中国听力语言康复科学杂志》 2017年第1期31-33,共3页 Chinese Scientific Journal of Hearing and Speech Rehabilitation
关键词 诺尔康 晨星人工耳蜗系统 ESRT Nurotron Morning-star cochlear system ESRT
  • 相关文献

参考文献3

二级参考文献24

  • 1陈雪清,王靓,孔颖,刘莎,刘博,莫玲燕,王硕,吴燕君,韩娜,孟黎晖,李海珍.用有意义听觉整合量表评估儿童人工耳蜗植入后听觉能力[J].中华耳鼻咽喉头颈外科杂志,2006,41(2):112-115. 被引量:39
  • 2杨彩虹,赵德安,陈雪清,刘博,李永新,郑军,韩德民.儿童人工耳蜗植入术后神经反应遥测阈值的变化[J].临床耳鼻咽喉头颈外科杂志,2007,21(13):598-600. 被引量:8
  • 3陈新 见:姜泗长 顾瑞 主编.声导抗测试[A].见:姜泗长,顾瑞,主编.临床听力学[C].北京:北京医科大学中国协和医科大学联合出版社,1999.191~232.
  • 4Jerger J,Fifer R,Jenkins H,et al.Stapedius reflex to electrical stimulation in a patient with a cochlear implant[J].Ann Otol Rhinol Laryngol, 1986,95:151.
  • 5Battmer RD,Laszig R,Lehnhardt E.Electrically elicited stapedius reflex in cochlear implant patients[J]. Ear Hear, 1990,11:370.
  • 6Todd NW,Ajayi E,Hasenstab MS,et al.Childhood otitis media and electrically elicited stapedius reflexes in adult cochlear implantees[J]. Otol Neurotol, 2003,24:621.
  • 7Van den Borne B,Srik AF,Mens LH,et al.Stapedius reflex measurements during surgery for cochlear implantation in children[J]. Am J Otol, 1996,17:554.
  • 8Stephan K,Welzl-Muller K.Post-operative stapedius reflex tests with simultaneous loudness scaling in patients supplied with cochlear implants[J]. Audiology, 2000,39:13.
  • 9Spivak LG,Chute PM.The relationship between electrical acoustic reflex thresholds and behavioral comfort levels in children and adult cochlear implant patients[J]. Ear Hear, 1994,15:184.
  • 10Cullington H.Preliminary neural response telemetry results[].British Journal of Audiology.2000

共引文献27

同被引文献15

引证文献5

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部