UPT-3A型上转发光免疫分析仪性能评价 被引量:1

The performance evaluation of UPT-3A immunoassay analyzer
下载PDF
导出
摘要 目的通过肌钙蛋白I(c Tn I)、N末端脑利钠肽前体(NT-pro BNP)以及降钙素原(PCT)定量检测对UPT-3A型上转发光免疫分析仪进行性能评价。方法收集临床血清标本,通过测定c Tn I、NT-pro BNP、PCT三个项目,分别进行精密度、准确度、线性范围分析及对三个项目进行生物参考区间验证。结果该分析仪的精密度、准确度、线性范围、参考区间均在仪器的允许范围之内。结论 UPT-3A型上转发光免疫分析仪检测结果准确可靠,精密度好,线性良好,符合相应的参考区间。主要性能基本符合质量目标要求,能够满足现有临床需要。 Objective By detection of the troponin I (cTnI) and N terminal pro brain natriuretic peptide (NTproBNP) and procalcitonin ( PCT) to evaluate the performance of type UPT3A chemiluminescence immunoassay analyzer. Methods The clinical serum samples were collected, and the precision, accuracy and linear range of the three items were determined by cTnI, NTproBNP and PCT, and the biological reference interval of the three items was tested. Results The precision, accuracy, linear range and reference interval of the analyzer were within the allowable range of the instrument. Conclusion The results of UPT3A type up conversion immunoassay analyzer are accurate and reliable, and the precision is good. The main performance is basically in accordance with the requirements of the quality objectives, and can meet the needs of the existing clinical.
机构地区 解放军
出处 《东南国防医药》 2016年第6期606-608,635,共4页 Military Medical Journal of Southeast China
关键词 上转发光技术(UPT) 精密度 准确度 线性范围 参考区间 性能评价 up-converting phosphor technology precision accuracy linearity reference interval performance evaluation