摘要
目的:研究不同植入扭矩(insertion torque,IT)及其种植体稳定系数(implant stability quotient,ISQ)值的种植体抵抗力作用下微动的能力,分析IT与ISQ能否反映种植体初期稳定性。方法:将24颗种植体以设定扭矩植入牛肋骨中并测量相应的ISQ值。安放好基台及全瓷冠后,进行水平向及垂直向加力,记录微动150μm时的力。结果:在每组样本中,IT和ISQ均与种植体初期稳定性不相关(P>0.05);在同一样本中,IT和ISQ可与种植体初期稳定性相关。结论:IT与ISQ反映种植体初期稳定性是有条件的;IT与ISQ不一定能反映种植体初期稳定性。
Objective:To analyze whether IT and ISQ can reflect dental implants primary stability.Methods:Twenty-four implants were inserted in 24 cow ribs,and the set torque and ISQ values were measured.After abutment and all-ceramic crown were placed,horizontal and vertical force were loaded until the micromotion reached150μm and the force values were recorded.Results:In each group,both IT and ISQ had no significant correlation with implants primary stability(P〉0.05).While in the same sample,IT and ISQ may have significant correlation with implant's primary stability.Conclusion:IT and ISQ can reflect dental implants primary stability under certain condition.
出处
《口腔医学研究》
CAS
北大核心
2017年第2期150-153,共4页
Journal of Oral Science Research
基金
2016年重庆高校创新团队建设计划资助项目(编号:CXTDG201602006)
重庆市高校市级口腔生物医学工程重点实验室资助项目
关键词
种植体
植入扭矩
种植体稳定系数
微动
初期稳定性
Implants
Insertion torque
Implant stability quotient
Micromotions
Primary stability