摘要
为进一步提高2016年国内比对中热释光剂量计测量的精密性和准确度,实验通过制定较为完善的实验方案,严格控制热释光剂量计的辐照、测读及退火条件,将热释光剂量计的重复性保持在1%以内;其刻度因子多次测量结果的误差也在1%以内;同时对其剂量响应和稳定性做了一定的研究。
In order to improve the precision and accuracy of thermo- luminescence dosimeters measurement in the 2016's domestic comparison, a comparatively consummate experimental program was made, in which the conditions while radiating, reading and annealing were strictly controlled. The results show that the repeatability of TLD is maintained within 1%, and the difference among repeated measurements of the calibration factor is also maintained within 1%. Meanwhile, some research was done about the dose response and the stability of TLD.
出处
《核电子学与探测技术》
CAS
北大核心
2016年第10期1029-1031,共3页
Nuclear Electronics & Detection Technology
基金
中国计量科学研究院基本科研业务(AKY1410)资助
关键词
热释光剂量计
剂量测量
影响因素
不确定度评定
thermo- luminescence dosimeters
dose measurement
influence factor
uncertainty evaluation