脊柱微创手术中医护人员遭受的X线辐射损害及应对措施 被引量:10

X-ray radiation damage of medical staff in minimally invasive spine surgery and its countermeasures
下载PDF
导出
摘要 目的探讨经皮内镜椎间盘切除伴椎间融合术中医护人员遭受的X线辐射的损害及应对措施。方法选择本科收治的行经皮内镜椎间盘切除伴椎间融合术的患者49例。术中X线照射采用飞利浦BV25型移动式C型臂X线机。术者手术过程中穿戴铅衣、铅围脖、含铅面罩、指环。分别检测术者身体不同部位和距辐射源不同距离的累计辐射剂量。结果防护设备外眼眶、左前胸、甲状腺、性腺的有效辐射剂量为(24.0±6.6)μSv、(96.5±15.2)μSv、(43.1±10.5)μSv、(82.4±20.6)μSv,防护设备内眼眶、左前胸、甲状腺、性腺的有效辐射剂量为(5.3±2.0)μSv、(3.5±4.7)μSv、(4.5±3.8)μSv、(7.5±2.4)μSv,防护设备内外的有效辐射剂量差异有统计学意义(P<0.05)。随着距辐射源距离的增加,累积辐射剂量显著降低,与1 m距离相比,2 m、3 m、4 m的累积辐射剂量均显著降低,差异有统计学意义(P<0.05)。结论脊柱微创手术中采取有效的措施防护医护人员遭受X线剂量辐射损害尤其重要,能够显著降低医护人员遭受X线辐射的剂量。 Objective To investigate X-ray radiation damage of medical staff in minimally invasive spine surgery and its countermeasures. Methods A total of 49 percutaneous endoscopic discectomy cases with interbody fusion in our hospital were selected,intraoperative X-ray radiation used PHILPS BV25 type mobile C-arm X-ray machine. The operator in the process wore lead clothes,lead collar,lead mask and rings. The cumulative radiation doses of different body parts and different radiation sources were detected. Results The radiation dose of outer eye socket,left chest,thyroid and gonad were( 24. 0 ± 6. 6) μSv,( 96. 5 ± 15. 2) μSv,( 43. 1 ± 10. 5) μSv,( 82. 4 ± 20. 6) μSv,respectively,and radiation dose of inner eye socket,left chest,thyroid and gonad were( 5. 3 ±2. 0) μSv,( 3. 5 ± 4. 7) μSv,( 4. 5 ± 3. 8) μSv,and( 7. 5 ± 2. 4) μSv. The difference was statistically significant in outside and inside radiation dosage of protective equipment( P 〈0. 05). The cumulative radiation dose was significantly decreased with the increase of distance from the source.Compared with 1 m distance,the cumulative radiation doses of 2 m,3 m and 4 m was significantly lower,and the difference was statistically significant( P〈 0. 05). Conclusion Effective measures to protect medical personnel from X-ray radiation damage in minimally invasive spine surgery is particularly important,it can significantly reduce the dose of X-ray radiation of medical staff.
作者 韦丹
出处 《实用临床医药杂志》 CAS 2017年第4期118-120,共3页 Journal of Clinical Medicine in Practice
关键词 微创手术 X线 辐射损害 防护 minimally invasive surgery X-ray radiation damage protection
  • 相关文献

参考文献3

二级参考文献47

  • 1刘鹏程,杜端明,陈在中,余宏建,戴鲁平,韩志华.介入放射治疗中的立体防护研究[J].中国辐射卫生,2006,15(1):18-20. 被引量:30
  • 2洪惠民,陈东杰,林辉.C形臂X射线机辐射防护状况调查[J].中国辐射卫生,2006,15(1):42-43. 被引量:15
  • 3CBZ130-2002,医用X射线诊断卫生防护标准[S].
  • 4张宏霞,张立新,高波,等.骨科手术室c型臂机对医护人员辐射剂量和防护策略研究[J].中国实用护理杂志,2012,28(增刊):73-75.
  • 5Lee K, Lee KM, Park MS, et al. Measurements of surgeons' exposure to ionizing radiation dose during intraoperative use of c-arm fluoroscopy[J]. Spine, 2012, 37(14): 1240-1244.
  • 6Choi HC. Fluoroscopic radiation exposure during percutaneous kyphoplasty[J]. J Korean Neurosurg Soc, 2011, 49(1): 37-42.
  • 7Taher F, Hughes AP, Sama AA, et al. 2013 young investiga- tor award winner: how safe is lateral lumbar interbody fusion for the surgeon? A prospective in vivo radiation exposure study[J]. Spine, 2013, 38(16): 1386-1392.
  • 8Payer M. "Minimally invasive" lumbar spine surgery: a criti- cal review[J]. Acta Neurochir(Wien), 2011, 153(7): 1455-1459.
  • 9Rodgers WB, Gerber EJ, Patterson J. Intraoperative and early postoperative complications in extreme lateral interbody fusion: an analysis of 600 cases[J]. Spine, 2011, 36(1): 26-32.
  • 10Anand N, Baron EM. Minimally invasive approaches for the correction of adult spinal deformity[J]. Eur Spine J, 2013, 22 (Suppl 2): 232-241.

共引文献34

同被引文献74

引证文献10

二级引证文献70

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部