4种不同眼压计测量值的比较 被引量:1

Comparison of Intraocular Pressure Values Measured by 4 Different Tonometers
下载PDF
导出
摘要 目的:比较分析非接触式眼压计(NCT)、Icare回弹式眼压计(RBT)、Tono-pen眼压计(TPT)和Goldmann眼压计(GAT)的眼压测量结果,评价不同眼压计在临床的应用价值。方法:收集我院临床患者36例(72眼)分别用NCT、RBT、TPT和GAT眼压计测量眼压,然后根据GAT测得的眼压值分为3组,A组(眼压≤21 mmHg)、B组(21 mmHg<眼压≤30 mmHg)、C组(眼压>30 mmHg),对比分析不同眼压程度分组下各种眼压计的测量结果。结果:NCT-GAT平均差值在A组为(0.22±0.43)mm Hg(t=1.936,P=0.075),B组为(0.90±1.26)mm Hg(t=2.156,P=0.063),C组为(7.94±3.02)mm Hg(t=6.958,P=0.000);RBT-GAT平均差值在A组为(0.55±0.57)mm Hg(t=3.656,P=0.003),B组为(0.05±0.45)mm Hg(t=0.347,P=0.738),C组为(-0.53±1.71)mm Hg(t=-0.825,P=0.441);TPT-GAT平均差值在A组为(-0.66±0.99)mm Hg(t=-2.531,P=0.026),B组为(-0.50±0.62)mm Hg(t=-2.434,P=0.041),C组为(-2.25±2.67)mm Hg(t=-2.223,P=0.068)。结论:NCT、RBT均适用于在临床工作中青光眼的诊断,因眼压≤21 mmHg时,RBT测量的眼压值与GAT相比差异有统计学意义,故最好不用于青光眼随访,TPT仅适用于青光眼筛查。 Objective: To compare and analyze the intraocular pressure (IOP) values measured by non-contact tonometer (NCT) , Ieare rebound tonometer (RBT) , Tono-pen tonometer (TPT) and Goldmann applanation tonometer (GAT) , and to evaluate the clinical application value of different tonometers in measuring IOP. Methods: IOP was measured by NCT, RBT, TPT and GAT tonometers in 36 patients (72 eyes) . The patients were divided into three groups according to IOP measured by GAT, group A (IOP≤21 mmHg), group B (21 mmHg〈IOP≤30 mmHg), group C (IOP〉30 mmHg) . Measurement results of different tonometers in different groups were compared and analyzed. Results: The average difference of NCT-GAT in group A was (0.22± 0.43) mmHg (t=1.936, P=0.075), (0.90±1.26) mmHg in group B (t=2.156, P=0.063), (7.94±3.02) mmHg in group C (t= 6.958, P=0.000) . The average difference of RBT-GAT in group A was (0.55±0.57) mmHg (t=3.656, P=0.003), (0.05±0.45) mmHg in group B (t=0.347, P=0.738), (-0.53±1.71) mmHg in group C (t=-0.825, P=0.441) . The average difference of TPT-GAT in group A was (-0.66±0.99) mmHg (t=-2.531, P=0.026), (-0.50+0.62) mmHg in group B (t=-2.434, P= 0.041) , (-2.25±2.67) mmHg in group C (t=-2.223, P=0.068) . Conclusions: Both NCT and RBT are suitable for the diagnosis of glaucoma in clinical practice. Because the measurement result with RBT is significantly different compared with GAT when IOP is ≤21 mmHg , it is not recommended for glaucoma follow-up. TPT is only suitable for glaucoma screening.
作者 李军 郭素平
出处 《沈阳医学院学报》 2017年第2期100-102,105,共4页 Journal of Shenyang Medical College
关键词 眼压 非接触式眼压计 ICARE回弹式眼压计 TONO-PEN眼压计 GOLDMANN眼压计 intraocularpressure non-eontacttonometer Icarereboundtonometer Tono-pentonometer Goldmanntonometer
  • 相关文献

参考文献5

二级参考文献58

  • 1高宗峰.安徽省桐城县青光眼流行病学调查[J].中华眼科杂志,1995,31(2):149-151. 被引量:26
  • 2胡铮,彭佑恩.北京市顺义县青光眼流行病学调查[J].中华眼科杂志,1989,25(2):115-119. 被引量:86
  • 3Pandav SS,Sharma A,Gupta A,et al.Reliability of ProTon and Goldmann applanation tonometers in normal and postkeratoplasty eyes[J].Ophthalmology,2002,109(5):979-984.
  • 4Garzoz HJ,Chung HS,Lang Y,et al.Intraocular pressure and photorefractive keratectomy a comparison of three different tonometers[J].Cornea,2001,20(1):33-36.
  • 5Duch S,Serra A,Castanera J,et al.Tonometry after laser in situ keratomileusis treatment[J].J Glaucoma,2001,10(4):261-265.
  • 6Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci,2003,44(9):3790-3794.
  • 7Munoz LA,Perez AL,Rahhal MS,et al.Assessment of applanation tonometry after hyperopic laser in situ keratomileusis[J].Cornea,2002,21(2):156-160.
  • 8Copt RP,Tomas R,Mermoud A.Corneal thickness and normal tension glaucoma[J].Arch Ophthalmol,1999,117(1):14-16.
  • 9Dekker PW,Robert YCA,Kanngiesser H,et al.Principles of contact lens tonometry[J].International Ophthalmology,1998,22(2):105-111.
  • 10Shah S,Chatterjee A,Mathai M,et al.Relationship between corneal thickness and messured intraocular pressure in a general opthalmology clinic[J].Ophthalmology,1999,106(11):2154-2160.

共引文献77

同被引文献15

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部