期刊文献+

流动注射在线光催化降解-分光光度法测定水中总磷 被引量:1

Determination of Total Phosphorus in Water by Flow Injection On-line Photocatalytic Digestion Spectrophotometry
下载PDF
导出
摘要 本文研究了利用二氧化钛光催化作用并添加适量过硫酸钾降解水体中的无机 /有机膦 ,通过流动注射分光光度法法测定总磷的含量。结果表明 :在二氧化钛和过硫酸钾的协同作用下 ,可有效降解无机 /有机膦。利用流动注射技术可以快速测定总磷的含量 ,在 30样 / h的采样频率下 ,检出限 (3σ)为 0 .0 2 mg/ L,线性范围为 0— 2 .5 mg/ L,RSD为 0 .2 1%。可用于工业循环水总磷的测定 。 A UV photocatalytic digestion system for the highly sensitive flow-inject spectrophotometric determination of total phosphorus in recycle water was described.Sample was digested in a UV photo-reactor with TiO 2 and potassium persulfate.A detection limit(3σ)of 0.02mg/L was achieved at a sampling frequency of 30h -1.The relative standard deviation was 0.21% for the determination of 0.4mg/L phosphorus (n=11).This method can be applied for the detection of total phosphorus in recycle water with satisfactory results.
出处 《光谱实验室》 CAS CSCD 2004年第6期1160-1162,共3页 Chinese Journal of Spectroscopy Laboratory
基金 国家自然科学基金资助项目 编号 :2 0 345 0 0 6
关键词 总磷 光催化降解 二氧化钛 流动注射分析 分光光度法 Total Phosphorus,Photocatalytic Digestion,Titanium oxide,Flow Injection Analysis,Spectrophotometry.
  • 相关文献

参考文献2

  • 1Kim Y C,Sasaki S,Kazuyoshi Y et al.Photocatalytic Sensor for the Determination of Chemical Oxygen Demand Using Flow Injection Analysis[].Analytica Chimica Acta.2001
  • 2Kim Y C,Sasaki S,Kazuyoshi Y et al.A Flow Method with Photocatalytic Oxidation of Disolved Organic Matter Using A SolidPhase(TiO2 ) Reactor Followed by Amperometric Detection of Consumed Oxygen[].Analytical Chemistry.2002

同被引文献922

引证文献1

二级引证文献15