摘要
对BG2015闪烁室测氡仪器在江西九江地震台2号井、甘肃平凉地震台安国井和云南省开远地震局滇18井不同井水条件进行实验,针对不同井条件下气氡测值的稳定性、连续性和可靠性进行比较分析,得到BG2015闪烁室测氡仪器应用于地震行业的可行性。
This paper aims to discuss the experiments on BG2015 measuring radon instrument in No.2 well of Jiujiang Seismic Station, Anguo well of Pingliang Seismic Station and No. 1 g well of Earthquake Administration of Kaiyuan City in Yunnan Province, which has different well water conditions. Comparative analysis is made according to stability, continuity and reliability of gas radon value in different well water conditions. The results show that BG2015 measuring radon instrument is feasible in earthquake industry.
出处
《防灾减灾学报》
2017年第1期44-51,共8页
Journal of Disaster Prevention And Reduction
基金
东华理工大学核技术应用教育部工程研究中心开放基金<水氡自动化装置研制>(HJSJYB2015-8)
江西省地震局新世纪优秀人才<气氡观测的水气比研究>(JXDZ-YXRC20163)
中国地震局三结合<氡无源传递标定方法与规程研究>(CEA-JC/3JH-161401)
关键词
BG2015闪烁室测氡仪
氡测值
对比分析
可行性
BG2015 scintillation chamber radon measuring instrument
radon value
comparative analysis