Echotracking联合microCT评价可降解支架植入后动脉弹性改变

Study of Artery Elasticity Changes of Rabbits' Abdominal Aorta after Biodegradable JDBM Stent Implanted by Echotracking Combined with Micro CT
下载PDF
导出
摘要 目的:通过建立可降解支架植入兔腹主动脉模型,探讨echotracking联合mircoCT在支架降解过程中动态观察动脉弹性改变的价值。方法:24只新西兰大白兔于DSA下在其腹主动脉随机植入一枚可降解镁合金支架(JDBM组,n=12)或镍钛合金支架(BMS组,n=12),分别于术前和植入术后即时、14天、30天、60天行echotracking获得支架处血管硬化参数β,并在术后第14天、30天、60天取材行micro CT评价支架降解情况。结果:(1)观察期间支架无移位或断裂,支架内血流通畅。(2)Micro CT提示BMS支架无明显变化,JDBM支架术后30天起已开始降解,观察期间支架基本保持原始结构,合金丝连续性完整。(3)术前及术后即时、第14天两组间β值无明显统计学差异(P>0.05),随植入时间的增加,JDBM组β值逐渐减小,BMS组β值逐渐增大;术后30天时JDBM组明显小于BMS组(P=0.036);术后60天时BMS组β值因支架处管壁无明显搏动无法获得。结论:Echotracking技术联合mirco CT可用于评价可降解镁合金支架植入后降解过程中血管弹性的改变情况。 Purpose: To determine whether echotracking combined with microCT could be used to assess the artery elasticity during the degradation processes. Methods: Either bare biodegradable magnesium alloy JDBM stent (JDBM group, n=12) or bare nitinol stent (BMS group, n=12) were implanted in New Zealand white rabbits' abdominal arteries. Echotracking was performed to obtain the stiffness parameter 13 before operation, immediate post-operation, 14, 30 and 60 days post-operation. The degradation of the stents was analyzed by mirco CT at the 14, 30 and 60 days after operation. Results:① There was no migration or fracture of all the stents,and the blood flow in stent was normal. ②MircoCT showed that the nitinol stents were with no changes. The JDBM stents began to degrade since 30 days afteroperation. But their structures maintained well and the struts were in continuity. ③There was no statistical significant difference of β between each groups before operation, immediate post-operation, or 14 days post-operation (P 〉 0.05). As the implanted time went by, β of JDBM group was significantly lower than that of BMS group(P =0.036) at 30 days post-operation. However, D of BMS group was immeasurable at the 60 days post-operation because of the undetectable vessel diameter changes. Conclusion: Echotracking combined with microCT could be used to assess the artery elasticity during the degradation processes of JDBM.
出处 《中国医学计算机成像杂志》 CSCD 北大核心 2017年第2期191-196,共6页 Chinese Computed Medical Imaging
关键词 可降解支架 Echotracking MICROCT 血管弹性 Biodegradable Stents Echotracking MicroCT Artery elasticity
  • 相关文献

参考文献3

二级参考文献20

  • 1WITTE F, KAESE V, HAFERKAMP H, SWITZER E, MEYER-LINDENBERG A, WIRTH C J, WINDHAGEN H. In vivo corrosion of four magnesium alloys and the associated bone response[J]. Biomaterials, 2005, 26: 3557-3563.
  • 2KANNAN M B, RAMAN R K. In vitro degradation and mechanical integrity of calcium-containing magnesium alloy inmodified-simulated body fluid[J]. Biomaterials, 2008, 29: 2306 2314.
  • 3WITTE F, KAESE V, HaAFERKAMP H, MEYER- L1NDENBERG A, WIRTH C J, WINDHAGEN H. In vivo corrosion of four magnesium alloys and the associated bone response[J]. Biomaterials, 2005, 26: 3557-3563.
  • 4FEYERABEND F, FISCHER J, HOLTZ J, WITTE F, WILLEMEIT R, DRUCHER H, VOGT C, HORT N. Evaluation of short-term effects of rare earth and other elements used in magnesium alloys on primary cells and cell lines[J]. Acta Biomaterialia, 2010, 6: 1834-1842.
  • 5Fischman DL, Leon MB, Bairn DS, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. Stent Restenosis Study Investigators. N Engl J Med, 1994,331: 496-501.
  • 6Dangas GD, Claessen BE, Caixeta A, et al. In-stent restenosis in the drug-eluting stent era. Journal of the American College of Cardiology, 2010, 56: 1897-907.
  • 7Kumbhani D J, Ingelmo CP, Schoenhagen E et al. Meta-analysis of diagnostic efficacy of 64-slice computed tomography in the evaluation of coronary in-stent restenosis. Am J Cardiol, 2009,103: 1675-1681.
  • 8Mehran R, Dangas G, Abizaid AS, et al. Angiographic patterns of in-stent restenosis: classification and implications for long-term outcome. Circulation ,1999, 100: 1872-1878.
  • 9Gebhard C, Fiechter M, Fuchs TA, et al. Coronary artery stents: influence of adaptive statistical iterative reconstruction on image quality using 64-HDCT. European heart journal cardiovascular Imaging. 2013; 14: 969-977.
  • 10Mahnken AH. CT Imaging of Coronary Stents: Past, Present, and Future. ISRN Cardiol, 2012, 2012: 139823.

共引文献110

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部