动脉血气分析检测ICU患者的电解质及阴离子间隙结果准确性探讨 被引量:7

The accuracy of arterial blood gas analysis in detecting electrolytes and anionic gap in ICU patients
原文传递
导出
摘要 目的 床旁动脉血气分析仪对ICU患者电解质及阴离子间隙结果与生化仪准确性比较.方法 回顾性分析376例ICU成年患者入ICU时的动脉血气分析仪和生化仪检测的血钠、钾、氯和碳酸氢根及计算的阴离子间隙(AG)结果.以生化仪检测结果为金标准,对两种检测方法进行比较.结果 生化仪检测的血钠、钾、氯及计算AG分别为121~183 mmol/L、2.13~6.77 mmol/L、86~146 mmol/L、1~62 mmol/L;动脉血气分析仪检测的血钠、钾、氯及计算AG分别为114~180 mmol/L、1.78~6.36 mmol/L、94~150 mmol/L、-7~40 mmol/L.动脉血气分析仪检测的血钠、钾、AG均低于生化仪结果,而血氯则高于生化仪.低、正常、高电解质及AG间两种检测方法的差值比较差异有统计学意义(P〈0.01).结论 动脉血气分析仪检测的电解质及AG结果与生化仪明显不同,动脉血气分析仪测量的电解质结果不能用于AG的计算. Objective To explore the accuracy of bedside arterial blood gas analyzer in detecting electrolytes and anionic gap (AG) in ICU patients, and compare it with auto-analyzer. Methods Results of blood sodium, potassium, chlorine, bicarbonate ions and AG measured by arterial blood gas analyzer and auto- analyzer of 376 ICU adult patients admitted to ICU were retrospectively analyzed. With the outcomes of auto-analyzer as standard, the correlation and difference of electrolytes and AG measured by 2 methods were analyzed. Results The sodium, potassium, chlorine and AG measured by auto-analyzer were 121 - 183 mmol/L, 2.13 - 6.77 mmol/L, 86 - 146 mmol/L and 1 - 62 mmol/L. The blood sodium, potassium, chlorine and AG measured by arterial blood gas analyzer were 114 - 180 mmol/L, 1.78 - 6.36 mmol/L, 94 - 150 mmol/L, -7 - 40 mmol/L. The blood sodium, potassium and AG measured by arterial blood gas analyzer were lower than those measured by auto-analyzer, but the blood chlorine was higher than that measured by auto- analyzer. There were statistical differences in the difference of low, normal and high electrolytes and AG between 2 methods (P〈0.01). Conclusions The electrolytes and AG measured by arterial blood gas analyzer and auto-analyzer are significantly different, and the electrolytes measured by arterial blood gas analyzer are unreliable to calculate AG.
出处 《中国医师进修杂志》 2017年第3期240-243,共4页 Chinese Journal of Postgraduates of Medicine
关键词 血气分析 电解质 阴离子 重症监护病房 回顾性研究 Blood gas analysis Electrolytes Anions Intensive care units Retrospective studies
  • 相关文献

参考文献2

二级参考文献10

  • 1The National Committee for Clinical Laboratory Standards Method Comparison and Bias Estimation Using Patient Samples: Approved Guideline [S]. NCCLS EP9-A 1995.
  • 2王明山.血气分析[M].北京:人民军队出版社,1995,36-98.
  • 3Adrogue HJ, Madias NE. Management of life-threatening acid-base disorders. First of two parts? [J].N Engl J Med, 1998, 338 (1): 26-34.
  • 4OH MS, Carrll HJ. The anion gap[J]. N Engl J Med, 1977, 297 (15): 814-817.
  • 5Gabow PA. Disorders associated with an altered anion gap [J]. Kidney Int, 1985, 27 (2) : 472-483.
  • 6Laffey JG, Kavanagh BP. Hypocapnia [J].N EnglJ Med, 2002, 347 (1) : 43-53.
  • 7Malhotra A. Low tidal volume ventilation in the acute respiratory distress syndrome [J]. N EnglJ Med, 2007, 357 (11): 1113- 1120.
  • 8Funk GC, Doberer D, Heinze G, et al. Changes of serum chloride and metabolic acid-base state in critical illness [ J ]. Anaesthesia, 2004, 59 (11): 1111-1115.
  • 9肖军,李小励.伴有三重酸碱失衡危重病人预后分析[J].急诊医学,1998,7(4):238-241. 被引量:6
  • 10任成山,陆海华,赵志强,郭中杰,高全杰,钱桂生.危重病患者的三重酸碱失衡分析[J].中国急救医学,2003,23(5):289-291. 被引量:6

共引文献7

同被引文献43

引证文献7

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部