摘要 All endoscopic imaging system using a plenoptic technique to reconstruct 3-D information is demonstrated and analyzed in this Letter. The proposed setup integrates a clinical surgical endoscope with a plenoptic camera to achieve a depth accuracy error of about 1 mm and a precision error of about 2 mm, within a 25 mm×25 mm field of view, operating at 11 frames per second. All endoscopic imaging system using a plenoptic technique to reconstruct 3-D information is demonstrated and analyzed in this Letter. The proposed setup integrates a clinical surgical endoscope with a plenoptic camera to achieve a depth accuracy error of about 1 mm and a precision error of about 2 mm, within a 25 mm×25 mm field of view, operating at 11 frames per second.
出处 《Chinese Optics Letters》 SCIE EI CAS CSCD 2017年第5期68-72,共5页 中国光学快报(英文版)
基金 supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Grant no.1R01EB020610