摘要 对RG-BS型测汞仪从日常工作状态、出现的故障、工作参数变化等方面进行阐述,并对仪器工作时抽气量的变化和捕汞管污染引起水汞观测数据的异常进行分析,为提高观测数据的连续性和稳定性、准确运用数据进行分析研究提供参考依据。 RG-BS mercury analyzer is expounded from the aspects of daily working state , failure occuparameters. The variation of the air quantity and the anomalies of water mercury observation data caused bymercury capture tube are analyzed also. This study provides the reference for improving the continuity and stability of the observation data and using the data accurately .
作者 张文男 杨静 黄春玲 ZHANG Wen-nan YANGJing HUANG Chun-ling(Xiaxian Central Seismological Station of Earthquake Administration of Shanxi Province, Xiaxian, Shanxi 044400, China State Key Observatory of Shanxi Rift System, Taiyuan, Shanxi 030025, China)
出处 《山西地震》 2017年第2期8-10,共3页 Earthquake Research in Shanxi
基金 山西省地震局科研项目(SBK-1211)部分内容
关键词 水汞观测 仪器故障 RG-BS型测汞仪 Water mercury observation Instrument failure RG-BS mercury analyzer