北京市部分医疗单位医用诊断X射线床边机性能分析

Analysis of Medical Diagnostic X-ray Bedside Mechanic Property in Some Medical Units in Beijing
下载PDF
导出
摘要 目的了解该市部分医疗单位医用诊断X射线床边机性能现状,为政府的有关决策提供参考。方法采用国家标准方法采样,利用国家标准判断结果是否合格。结果床边机合格率仅为75.00%。其中,"光野与照射野四边的偏离"合格率最低(85.42%),其次是"SID值的偏离"(87.50%),第三是"管电压指示的偏离"(89.58%)。结论该市部分医疗单位医用诊断X射线床边机需要更新及维修。 Objective To know the medical diagnostic x-ray bedside mechanic property in some medical units in Beijing and provide reference for the related decision of the government.Methods The national standardized method was used for sampling and whether the results were qualified or not was determined by the national standards.Results The qualified rate of bedside mechanism was only 75.00%,the qualified rate of light field and irradiation field quadrilateral deviation was the lowest(85.42%),the deviation of SID was the second(87.50%) and the deviation of tube voltage guidance was the third(89.58%).Conclusion The medical diagnostic x-ray bedside machines in some medical units in our city need update and maintenance.
出处 《中国卫生产业》 2017年第14期60-61,共2页 China Health Industry
关键词 X射线床边机 性能 现状 X-ray bedside machine Property Status
  • 相关文献

参考文献5

二级参考文献29

  • 1王强,朱益民,舒丽萍,廖云华.车载X射线机防护情况调查[J].浙江预防医学,2009,21(2):41-42. 被引量:3
  • 2陈以水,王琦,李南翔,熊晓英,邓磊.江西省医用诊断X射线机性能现状调查分析[J].中华放射医学与防护杂志,2011,31(6). 被引量:8
  • 3GB/T17589-1998.X射线计算机断层摄影装置影像质量保证检测规范[S].1998.
  • 4WS/T189-1999.医用X射线诊断设备影像质量控制检测规范[S].1999.
  • 5卫生部43号令.大型医用设备配置与应用管理暂行办法[S].2005.
  • 6卫生部卫监发[1998]第18号文件.X射线计算机体层摄影装置(CT)应用质量检测与评审规范[S].1998.
  • 7国家质量监督检验检疫总局.GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2003.
  • 8国家质量监督检验检疫总局.GB18871-2002电离辐射防护与辐射源安全基本标准.北京:中国标准出版社,2003.
  • 9GB18871-2002,电离辐射防护与辐射源安全基本标准[S].2002.
  • 10中华人民共和国卫生部.放射诊疗管理规定,2006,01:24

共引文献26

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部