期刊文献+

一种真耳-耦合腔差值(RECD)测量方法的有效性研究

Validity Study of a New RECD Measurement Method
下载PDF
导出
摘要 目的考察一种新的真耳-耦合腔差值(real-ear-to-coupler difference,RECD)测量方法——探管固定同步介入法(tube fixation and synchronous insertion method,TFSI)的可行性和有效性。方法对20例(40耳)受试者(成人和儿童各10例)分别应用传统方法和TFSI方法进行RECD测量,考察两种方法测量结果的一致性;对13例受试者(26耳)运用TFSI方法进行两次RECD测量,考察两次测量结果的一致性。结果采用传统方法和TFSI方法测得的RECD值差异不显著(P〉0.05);采用TFSI方法先后两次测得的RECD值差异不显著(P〉0.05);TFSI方法耗时(平均2.15分钟)明显少于传统方法(平均6.03分钟)(P〈0.05)。结论TFSI方法可以准确、有效测量RECD,其耗时较短、可重复性好。 Objective To test the vadility of a new RECD measuring method,Tube Fixation and Synchronous Insertion Method(TFSI). Methods Twenty subjects(40 ears) were measured with the conventional RECD measuring method and TFSI method respectively. Thirteen subjects(26 ears) were tested twice with TFSI method. Results The two sets of RECD values from two methods were shown no significant difference (P〉0.05). The RECD values twice with TFSI method were compared and proved to have no significant difference (P〉0.05). The operation time of using TFSI method(2.15 min) was significantly less than that of using the conventional method(6.03 min)(P〈0.05). Conclusion The new RECD measuring method was proved to be effective, time-saving and repeatable.
出处 《中国听力语言康复科学杂志》 2017年第4期291-295,共5页 Chinese Scientific Journal of Hearing and Speech Rehabilitation
关键词 真耳-耦合腔差值 探管固定同步介入法 真耳测试 探管麦克风测量 助听器验配 Real-ear-to-coupler difference (RECD) Tube fixation and synchronous insertion method (TFSI) Real ear measurement (REM) Probe microphone measurement (PMM) Hearing aid fitting
  • 相关文献

参考文献1

二级参考文献14

  • 1Bentler RA, Paviovic CV. Transfer functions and correction factors used in hearing aid evaluation and research[J]. Ear and Hearing, 1989,10 :58.
  • 2Feigin JA, Kopun JG, Stelmachowicz PG,et al. Probe-tube microphone measures of ear-canal sound pressure levels in infants and children[J]. Ear and Hearing, 1989, 10:254.
  • 3Nelson Barlow NL, Auslander MC, Rines D, et al. Probe-tube microphone measures in hearing impaired children[J]. Ear and hearing,1988,9:243.
  • 4Moodie KS, Seewald RC, Sinclair ST. Procedure for predicting real-ear hearing aid performance in young children[J]. American Journal of Audiology,1994,3:23.
  • 5Zelisko DL,Seewald RC,Gagn JP. Signal delivery/real ear measurement system for hearing aid secletion and fitting[J]. Ear and Hearing, 1992,13: 460.
  • 6Fikret-Pass S, Revit LJ. Individualized correction factors in the preselection of hearing aids[J].Journal of Speech and Hearing Research, 1992,35:384.
  • 7Seewald RC,Hudson SP,Gagné JP, et al. Comparison of two methods for estimating the sensation level of amplified speech[J]. Ear and Hearing,1992,13:142.
  • 8Seewald RC, Ross M, Stelmachowicz PG. Selecting and verifying hearing aid performance characteristics for young children[J]. Journal of the Academy of Rehabilitative Audiology, 1987, 20:25.
  • 9Seewald RC, Moodie KS, Sinclair ST, et al. Predictive validity of a procedure for pediatric hearing instrument fitting[J]. American Journal of Audiology, 1999, 8: 143.
  • 10Westwood GF,Bamford JM. Probe-tube microphone measures with very young infants: real ear to coupler difference and longitudinal changes in real ear unaided response[J]. Ear and Hearing, 1995, 16:263.