期刊文献+

石墨炉原子吸收法测定土壤中镉的不确定度评定 被引量:3

Evaluation on Uncertainty of Cadmium in Soil Determined by Graphite Furnace Atomic Absorption
下载PDF
导出
摘要 评定了石墨炉原子吸收法测定土壤样品中镉的测量不确定度。在测定过程中,对样品称量、水分测定、样品消解、样品定容、标准溶液配制、标准曲线拟合等影响不确定度的分量进行了评定。按照《测量不确定度评定与表示》(JJF 1059.1—2012)的规定进行合成,给出了扩展不确定度,镉含量结果的表示式为(0.115±0.015)mg/kg,k=2。 The uncertainty was estimated for determining cadmium in soil by graphite furnace atomic absorption. Uncertainty components such as sample weighing,moisture determination,process of digestion,volume,preparation of standard solutions,curve fitting were evaluated and synthesized according to Evaluation and Expression of Uncertainty in Measurement( JJF 1059. 1—2012). The expanded uncertainty was given. The mathematical expression of analysis of cadmium was given as( 0. 115 ± 0. 015)mg/kg,k = 2.
出处 《环境监控与预警》 2017年第4期22-25,共4页 Environmental Monitoring and Forewarning
关键词 石墨炉原子吸收分光光度法 不确定度 土壤 Graphite furnace atomic absorption Cadmium Uncertainty Soil
  • 相关文献

参考文献6

二级参考文献45

  • 1刘国伟,尹洪宗,何锡文.不确定度评定中离群值的检验及计算机编程[J].冶金分析,2004,24(4):63-66. 被引量:14
  • 2北京市环境保护监测中心.环境监测测量不确定度评定[M].北京:中国计量出版社,2009,1.
  • 3国家质量技术监督局.JJF 1135-2005化学分析测量不确定度评定[S].2005.
  • 4环境保护部,HJ583-2010,环境空气苯系物的测定固体吸附/热脱附-气相色谱法[S].北京:中国环境科学出版社,2010.
  • 5GB/T27407-2010.实验室质量控制利用统计质量保证和控制图技术评价分析测量系统的性能[S].中国国家标准化管理委员会,2010.
  • 6国家质量监督检验检疫总局.JJF 1059.1-2012测量不确定度评定与表示[S].北京:中国标准出版社,2013.
  • 7HJ/T 166-2004.土壤环境监测技术规范[S],2004.
  • 8CNAS-GL05(2006)测量不确定度要求的实施指南[S].
  • 9CNAS-GL06(2006)化学分析中不确定度的评估指南[S].
  • 10EPA TO - 14A - 1999 Determination of volatile organic com- pounds (VOCs) in ambient air using specially prepared canisters with subsequent analysis by gas chromatographyl S 1-.