摘要 目的:用化学发光微粒免疫分析(CMIA)法测定血清或血浆高敏肌钙蛋白(hsTnI),从而了解其检测性能。方法:用雅培公司的ARCHITECT系统i2000SR化学发光分析仪,使用配套的试剂、校准品和质控品,参照SOP文件对该方法的精密度、准确度、线性、携带污染率、灵敏度进行评价,并验证参考值区间。结果:(1)CMIA法测定hsTnI的批内和批间精密度,双水平的CV值均小于规定的范围,即批内、批间精密度的CV值均小于7.5%。(2)hsTnI的准确度测定值与靶值的偏倚在0.25%~6.31%之间,均低于EQA允许误差的30%。(3)hsTnI线性范围0.0-50000.0 pg/m L。(4)携带污染率实验中,hsTnI表现出良好的抗污染能力。(5)ARCHITECT系统i2000SR化学发光仪对hsTnI检测的分析灵敏度(最低检测限)均在厂家提供的范围之内;hsTnI的参考值区间的验证,男女均符合厂家提供的参考范围。结论:CMIA法测定hsTnI的结果评价其精密度和准确度、线性、携带污染率、灵敏度,均表现出良好的性能结果,满足临床要求。
作者 杨佳
出处 《淮海医药》 CAS 2017年第5期563-564,共2页 Journal of Huaihai Medicine
  • 相关文献

参考文献1

二级参考文献14

  • 1U.S. Department of Health and Human Services. 42 CFR Part 493: medicare, medicaid and CLIA programs;Laboratory requirements relating to quality systems and certain personnel qualifications; Final rule[ J ]. Federal Register, 2003,68:3640-3714.
  • 2Jan SK. Evaluation of assay systems [ J ]. Clin Lab News,2001,27 : 10-14.
  • 3Laessig RH, Ehrmeyer SS, Leinweber JE. Intralaboratory performance requirements necessary to pass proficiency testing: CAP-1990 vs CLIA-1967 ( March 14,1990) formats compared [ J ]. Clin Chem, 1992,38 :895-903.
  • 4U.S. Department of Health and Human Services.Medical devices;Current good manufacturing practice(CGMP) final rule :quality system regulation. 21 CFRParts 808,812 and 820 [ J ]. Federal Register, 1996,61 : 52602-52662.
  • 5National Committee for Clinical Laboratory Standards.Evaluation of precision performance of clinical chemistry devices, approved guideline [ S ]. Wayne, PA:EPSoA. NCCLS. 1999.
  • 6U.S. Department of Health and Human Services.Medicare,medicaid and CLIA programs; Regulationsimplementing the clinical laboratory improvement amendments of 1998 (CLIA). Final rule [ J ]. Fed Regist, 1992,57:7002-7186.
  • 7VITROS system verification [ S ]. Ortho-Clinical Diagnotics, USA ,2004.
  • 8National Committee for Clinical Laboratory Standards.User demonstration of performance for precision and accuracy [ S ]. Wayne, PA : EP15-P. NCCLS. 1998.
  • 9National Committee for Clinical Laboratory Standards.Method comparison and bias estimation using patientsamples, approved guideline [ S ].2nd ed. Wayne, PA :EP9-A2. NCCLS. 2002.
  • 10National Committee for Clinical Laboratory Standards.Evaluation of the linearity of quantitative measurement procedures : a statistical approach, approved guideline[S]. Wayne,PA:EP6-A. NCCLS. 2003.