期刊文献+

PCM控制及自动进样技术改进非甲烷总烃测定气相色谱仪 被引量:1

GC Determination of Non-methane Hydrocarbons has been Improved by Using Auto Injection and PCM Technique
下载PDF
导出
摘要 本文介绍了一种改进的非甲烷总烃测定气相色谱仪及其在环境监测中的应用。改进型的非甲烷总烃测定气相色谱仪采用PCM载气分流控制及自动进样技术相结合的方法促使样品经自动进样器自动取样至定量环后由PCM分流控制的两路载气流分别直接推送至总烃柱和甲烷柱后又流经同一微流平板进入单一FID检测器,利用保留时间的不同先后对样品中总烃和甲烷进行检测。方案中,色谱仪本身无需再配置进样口,只需单一检测器,在提高检测分析自动化程度和效率的同时,又节约成本,减少能耗,满足非甲烷总烃的检测要求。 This study presents An improvement in gas chromatography(GC)determination of non-methane hydrocarbons(NMHC)using auto injection and PCM technique. The improved GC with PCM carrier-gas controlling system,auto injection sampling system,non-injector,dual-column,micro flow plate and single FID detector determine NMHC automatically. With this improvement,the determination of NMHC is fast,easy,real cost saving,time and effort,and automation and analysis efficiency could be greatly improved.
作者 许百环 Xu Bai-huan(Quanzhou Environmental Monitoring station, Quanzhou,Fujian 362000, Chin)
出处 《福建分析测试》 CAS 2017年第5期6-11,共6页 Fujian Analysis & Testing
关键词 PCM 自动进样 微流平板 气相色谱 非甲烷总烃 PCM Auto injection Micro flow plate GC Non-methane hydrocarbons
  • 相关文献

参考文献2

二级参考文献8

  • 1陆凤家,曹玲秋.总烃和非甲烷烃的气相色谱法测定[J].上海环境科学,1993,12(8):29-31. 被引量:4
  • 2日本化学会 李芥春 等.碳氢化合物污染及其对策[M].北京:科学出版社,1987,6..
  • 3HJT38-1999,固定污染源排气中非甲烷总烃的测定气相色谱法[S].
  • 4陆凤家,上海环境科学,1993年,12卷,8期,287页
  • 5国家环境保护局,空气和废气监测分析方法,1990年,147页
  • 6日本化学会,碳氢化合物污染及其对策,1987年,1页
  • 7国家环境保护总局《空气和废气监测分析方法》编委会.空气和废气监测分析方法(第四版)[M].北京:中国环境科学出版社,2003-09.187-190.
  • 8齐文启,武攀峰,敬红.关于非甲烷总烃(NMHC)的测定及结果表示[J].中国环境监测,2009,25(4):30-31. 被引量:18

共引文献24

同被引文献9

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...