期刊文献+

基于频率锁定吸收光谱技术的变压器故障特征气体检测研究 被引量:4

Detection Study of Transformer Fault Characteristic Gases Based on Frequency-locking Absorption Spectroscopy Technology
下载PDF
导出
摘要 高准确度、高灵敏度检测故障特征气体是基于油中溶解气体分析的变压器早期潜伏性故障诊断的关键。该文开展了基于频率锁定吸收光谱技术的变压器故障特征气体检测研究。基于电四极跃迁理论,论证了H_2等同核双原子分子同样存在吸收效应;为提高H_2最小浓度的检测极限,搭建了频率锁定腔增强吸收光谱气体检测平台,首次实现了基于吸收光谱技术的变压器故障特征气体H_2的定性与定量检测分析,检测准确度与重复性分别达到约93.82%及94.35%。腔内气体压强为101k Pa时,H_2最小检测浓度的实验值约为7.5μL/L,与计算值的相对偏差仅为3.8%。此外,基于激光器阵列技术,实现了其他6种变压器故障特征混合气体CO、CO2、CH_4、C_2H_2、C_2H_4与C_2H_6的吸收光谱检测,其检测准确度与最小检测浓度同样满足变压器油中溶解气体分析的需求。该研究为研制基于频率锁定吸收光谱技术的变压器故障特征气体现场检测系统提供了理论依据和技术支撑。 Highly accurate and sensitive detection of fault characteristic gases is the key to diagnosing incipient transformer faults through dissolved gas analysis. In this paper, the detection study of transformer fault characteristic gases based on the frequency-locking absorption spectroscopy technology was carried out. Based on electric quadruple transitions theory, it had been proved that homonuclear diatomic molecule, such as H2, also exist absorption effect. To improve the detection limit of minimum concentration of H2, a frequency-locking cavity enhanced absorption spectroscopy platform for gas measurement was built up. The qualitative and quantitative analysis of fault characteristic gas H2 in transformer was achieved for the first time, the detection accuracy and repeatability is about 93.82% and 94.35%, respectively. The experimental value of minimum detectable concentration of H2 reaches approximately 7.5μL/L, the relative deviation of is only 3.8% when comparing to the calculated value. In addition, based on laser array technology, the measurement of other 6 kinds of fault characteristic gas-mixtures (CO、 CO2、 CH4、 C2H2、 C2H4 and C2H6) was achieved, the detection accuracy and minimum detectable concentration of which satisfy the requirement of dissotved gases analysis in transformer oil. All of which provides a theoretical basis and technical support for develop a live detection system for transformer fault characteristic gas based on frequency-locking absorption spectroscopy technology.
出处 《中国电机工程学报》 EI CSCD 北大核心 2017年第18期5504-5510,共7页 Proceedings of the CSEE
基金 国家自然科学基金项目(61605020) 重庆市基础科学与前沿技术研究专项(cstc2016jcyj A0427) 中国及重庆市博士后科学基金项目(2016M600722 2017T100675 Xm201602)~~
关键词 油中溶解气体分析 H2吸收谱线 电四极跃迁 频率锁定增强腔 激光器阵列 oil-dissolved gas analysis absorption lines of H2 electric quadruple transitions frequency-locking enhanced cavity laser array
  • 相关文献

参考文献5

二级参考文献96

共引文献108

同被引文献74

引证文献4

二级引证文献65

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部