期刊文献+

气质联用法测定膜式氧合器中偏苯三酸三辛酯的溶出量 被引量:1

Determination of Tri-Octyltri-Meta-Benzoate(TOTM) Released from Membrane Oxyenator Oxyenator Using Gas Chromatography-Mass Spectrometry
下载PDF
导出
摘要 建立一种气相色谱-质谱联用仪测定膜式氧合器中偏苯三酸三辛酯(TOTM)的溶出量方法,采用70%乙醇水混合液作为浸提介质,模拟临床使用。研究结果表明,该方法在0.125~20μg/ml范围内线性关系良好,线性关系为r=0.9963,TOTM的回收率为80.89%~104.81%,RSD小于5.00%(n=3),说明该方法准确度高,重复性好。本研究还通过使用10%脂肪乳作为浸提介质,对比集成式膜式氧合器在两种浸提介质中TOTM溶出量随时间的变化情况,结果表明10%脂肪乳浸提液中TOTM溶出量几乎不随时间变化,且在允许限量范围内。 Gas chromatography-mass spectrometry(GC-MS)was established to determinate tri-octyltri-meta-benzoate(TOTM)released from membrane oxyenator.Aqueous ethanol solution was the extraction medium in simulated clinical use.The experimental results showed good linearity in the range of 0.125~20μg/ml and the correlation coefficient was 0.9963.The spiked recoveries of TOTM ranged from 80.89%-104.81%,and the relatives standard deviations were less than 5.00%(n=3).The results showed that method was accurate and reproducible.The 10%fat emulsion also was used as extraction medium in this paper,and the TOTM dissolution changes with time were compared between the two extraction mediums.Finally found that dissolving-out amount of TOTM almost had no change with time in 10%fat emulsion,and was in the allowable limit range.
出处 《医疗装备》 2017年第19期68-70,共3页 Medical Equipment
关键词 气相色谱-质谱联用仪 氧合器 TOTM 溶出量 GC-MS Membrane oxyenator TOTM Dissolving-out amount
  • 相关文献