ABL800血气分析仪与BC5800血细胞分析仪检测HGB与HCT结果的比较研究 被引量:7

Comparative study on HGB and HCT results by ABL800 blood gas analyzer and BC5800 blood cell analyzer
下载PDF
导出
摘要 目的:研究ABL800血气分析仪与BC5800血细胞分析仪测量血红蛋白(HGB)与红细胞压积(HCT)结果的差异。方法:对2016年7月17—21日的97例肝素抗凝动脉血与EDTA-3K抗凝静脉血标本及8月4—9日的63例肝素抗凝动脉血标本使用ABL800血气分析仪与BC5800血细胞分析仪分别同时测量HGB及HCT。结果:ABL800血气分析仪测定的HGB与HCT大部分高于BC5800血细胞分析仪测定结果。结论:ABL800血气分析仪检测HGB与HCT与传统方法明显不同,检测结果应以血细胞分析仪结果为准,血气分析仪结果作为参考。 Objective To explore the difference between hemoglobin(HGB) and hematocrit(HCT) results by ABL800 blood gas analyzer and BC5800 blood cell analyzer respectively. Methods ABL800 blood gas analyzer and BC5800 blood cell analyzer were used to detect HGB and HCT in 97 samples of heparin-anticoagulated arterial blood and EDTA-3K anticoagulated versus blood from July 17 to 21, 2016 as well as 63 samples of heparin-anticoagulated arterial blood from August 4 to 9, 2016. Results Almost all the results by ABL800 blood gas analyzer were higher than those by BC5800 blood cell analyzer. Conclusion There're significant differences between ABL800 blood gas analyzer and traditional way when used to detect HGB and HCT. Blood cell analyzer is the preferred choice to detect HGB and HCT, and the results by the blood gas analyzer can be used for references.
机构地区 解放军
出处 《医疗卫生装备》 CAS 2017年第7期97-100,共4页 Chinese Medical Equipment Journal
关键词 血气分析 血细胞分析 血红蛋白 红细胞压积 血标本 analysis of blood gas analysis of blood cell HGB HCT blood sample