摘要 目的 :建立能谱CT单能量CT值与组织密度的关系,试图通过测量CT值来间接计算甲状腺的密度,为测量甲状腺质量提供基础。方法:选取不同成分的固体与液体、液体与液体的混合液、实体甲状腺组织,分别测量CT值与其密度,分析两者的关系。结果:自然界中,物质的密度与其CT值没有相关关系;当组织由相对固定的成分组成且每种成分含量不同时,其CT值与组织中大原子序数的物质含量直接相关;甲状腺组织的CT值与其密度及其内的碘含量直接相关。结论:甲状腺的能谱CT值与其密度之间具备相关关系,可以用能谱CT来间接测量甲状腺组织的密度。 Objective To investigate the relationship between the single energy CT value and the tissue density and further evaluate the mass of thyroid. Methods Different substances respectively composed of liquid and solid, liquid and liquid,thyroid gland were selected, and the relationship was investigated between the single energy CT value and the tissue density.Results The CT value was closely related to the content of high atomic number substance. The CT value of thyroid was closely related to the content of iodine. Conclusion It is feasible to measure the thyroid density by the energy spectrum CT.
出处 《医疗卫生装备》 CAS 2017年第10期89-92,共4页 Chinese Medical Equipment Journal
关键词 能谱CT 甲状腺 密度 CT值 甲亢 energy spectrum CT thyroid density CT value hyperthyroidism
  • 相关文献

参考文献7

二级参考文献42

  • 1王懋基,蔡鑫,涂承林.中国重力勘探的发展与展望[J].地球物理学报,1997,40(S1):292-298. 被引量:24
  • 2陈再君,蒋宁一,陶滔,李春艳,桂文来,罗茂香,赵小艳.用SPECT断层显像体积积分法测定甲状腺质量[J].中华核医学杂志,2004,24(3):172-173. 被引量:10
  • 3Johannes W,John MH,Peter P,et al.Comparison of methods for thyroid volume estimation in patients with Graves'disease.Eur J Nucl Med Mol Imaging,2003,30(4):525-531.
  • 4Lyshchik A,Drozd V,Reiners C.Accuracy of three-dimensional ultrasound for thyroid volume measurement in children and adolescents.Thyroid,2004,14(2):113-120.
  • 5Edmond NG,Teresa CHEN,Rashida LAM,et al.Three-dimesional ultrasound measurement of thyroid volume in asymptomatic male Chinese.Ultrasound in Medieine & Biology,2004,30(11):1427-1433.
  • 6Ying M,Sin MH,Pang SF.Sonographic measurement of thyroid gland volume:a comparison of 2D and 3D ultrasound.Radiography,2005,11(4):242-248.
  • 7Rago T,Bencivelli W,Scutari M,et al.The newly developed three-dimensional (3D) and two-dimensional (2D) thyroid ultrasound are strongly correlated,but 2D overestimates thyroid volume in the presence of nodules.J Endocrinol Invest,2006,29(5):423-426.
  • 8Pang BS,Kot BC,Ying M.Three-dimensional ultrasound volumetric measurements:is the largest number of image planes necessary for outlining the region-of-interest? Ultrasound Med Biol,2006,32(8):1193-1202.
  • 9Pang SF,Sin MH,Ying M.Evaluation of the accuracy and reliability of three-dimensional ultrasound in volumetric measurements using regular and irregular-shaped tissue phantoms.The Hong Kong Radiographers J,2005,8(1):17-20.
  • 10Peeters EY,Shabana WM,Verbeek PA,et al.Use of a curved-array transducer to reduce interobserver variation in sonographic measurement of thyroid volume in healthy adults.J Clin Ultrasound,2003,31(4):189-193.

共引文献121

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部