摘要 全身计数器用于对人体内照射污染的监测,对最小探测活度(MDA)有较高的要求。为使全身计数器具备较好的灵敏度,需要降低本底辐射和提高射线探测效率。基于MCNP5模拟了全身计数器中NaI探测器在不同几何位置时的能量响应,对探测器位置及准直器角度进行研究,同时对不同厚度的屏蔽结构进行了仿真计算分析。研究表明,当屏蔽钢板厚度为14.9cm时,能够较大程度地降低本底辐射。在源探结构上探测腔准直器角度在26°以上,探测器与体模的距离为20cm时,能够得到理想的能谱响应和探测效率。 The whole body counter is used to monitor the contamination of the human body and has a high requirement for the minimum detection limit(MDA).In order to have a better sensitivity for the whole body counter,it is necessary to reduce the background radiation and improve the radiation detection efficiency.Based on MCNP5,the energy response of the NaI detector in the geometric position is simulated.The position of the detector and the angle of the collimator are studied.At the same time,the shielded structure with different thickness is simulated and analyzed.The study shows that when the thickness of the shielded steel plate is 14.9 cm,the background radiation can be reduced to a large extent.In the source structure to detect the cavity collimator angle above 26°,when the distance between the detector and the body mold is 20 cm,the detector can get the ideal energy spectrum response and detection efficiency.
出处 《强激光与粒子束》 EI CAS CSCD 北大核心 2017年第12期94-97,共4页 High Power Laser and Particle Beams
关键词 MCNP模拟 NaI探测器 探测效率 本底 脉冲沉积 MCNP simulation NaI detector detection efficiency background pulse deposition
  • 相关文献

参考文献3

二级参考文献19

  • 1王成焘.中国力学虚拟人[J].医用生物力学,2006,21(3):172-178. 被引量:37
  • 2李丹,赖万昌,王广西,郭伟.蒙特卡罗方法计算介质对窄束γ射线的质量衰减系数[J].核电子学与探测技术,2007,27(2):415-417. 被引量:7
  • 3章晔,华荣洲,石柏惧.放射性勘查方法[M].北京:原子能出版社,1990.
  • 4Tung - Chai Ling, Chi - Sun Poon, Wai - Shung Lain, et al. Utilization of recycled cathode ray tubes glass in cement mortar for X - ray radiation - shielding applications [ J ]. Journal of Hazardous Materials, 2012, 199 - 200:321 - 327.
  • 5Meat F, Yot P, Cambon M, et al. The Characteriza- tion of Waste Cathode - ray Tube Glass [J]. Waste Management, 2006, 26(12) : 1468 - 1476.
  • 6E. Yilmaz, H. Baltas , E. Kiris ,et al. Gama ray and neutron shielding properties of some concrete ma- terials [ J ]. Annals of Nuclear Energy , 2011,38 (10) :2204 -2212.
  • 7许淑艳,张贵山.NaI(T1)晶体对1射线响应函数的蒙特卡罗计算[J].原子能科学技术,1982,2期:240-247.
  • 8田家云,张文仲.GM计数管内光子能量沉积影响特性研究[C].第十三届全国核电子学与核探测技术学术年会,2006.
  • 9ICRP. Conversion coefficients for use in radio - logi- cal protcctlon against external radiation, ICRP Publlca- tion 74 [R ]. Oxford: Pergamon Press,1996.
  • 10国家质量监督检验检疫总局.JJG393-2003辐射防护用x、γ辐射剂量当量(率)仪和监测仪检定规程[S],2003.