摘要
目的探讨软组织高频焊接设备闭合血管的可行性。方法以家犬为实验对象,分离不同直径的动脉、静脉,随机分为两组,分别采用两种软组织高频焊接设备EKVZ-300 PATONMED和NERC-300闭合血管,并检测闭合端血管的爆破压。结果 EKVZ-300、NERC-300两种高频焊接设备均能有效完成不同直径家犬动脉和静脉的闭合,且闭合后爆破压在两组间无明显差异(P>0.05)。结论 NERC-300能够完成离体家犬动脉和静脉的闭合,其闭合效力等同于EKVZ-300。
Objective To investigate the feasibility of high-frequency live tissue welding devices in vascular closure. Methods The artery and vein with different diameters of domestic dogs were separated and closed using high-frequency live tissue welding devices (EKVZ-300 and NERC-300) respectively, and the bursting pressure was compared between two devices. Results Both EKVZ-300 and NERC-300 could close the blood vessels effectually, and the bursting pressure showed no statistical difference between two devices (P〉0.05). Conclusions NERC-300 can close the in-vitro artery and vein of domestic dons. and the effectiveness is similar to EKVZ-300)
出处
《临床医学工程》
2017年第12期1673-1674,共2页
Clinical Medicine & Engineering
基金
广州市科技计划项目(201508030035)
广东省战略性新兴产业核心技术攻关项目(2011A081402004)
广东省科技计划项目(2017A010103009)
关键词
软组织高频焊接设备
血管闭合
High-frequency live tissue welding device
Vascular closure