全自动血凝仪CA-7000性能及生物参考区间的验证与评价 被引量:6

Verifying and evaluating automated blood coagulation analyzer CA-7000:its performance and biological reference interval
下载PDF
导出
摘要 目的了解攀钢集团总医院检验科实验室新购SYSMEX全自动血凝仪CA-7000的性能指标是否符合相关要求以及上海太阳试剂厂商提供的参考区间能否直接应用于临床。方法(1)批内不精密度:取新鲜混合血浆[凝血酶原时间(PT)、活化部分凝血活酶时间(APTT)、纤维蛋白原(FIB)为3个浓度水平,凝血酶时间(TT)、D-二聚体、纤维蛋白原降解产物(FDP)为正常浓度水平]连续测定各20次,计算标准偏差(SD)、变异系数(CV)。(2)日间不精密度:采用两个水平的质控品,每天测定1次,连续20 d,计算SD、CV。(3)携带污染率:测定稀释的血浆3次后,测定原浓度血浆3次,再测定稀释的血浆3次,计算携带污染率。(4)线性范围:测定5个浓度的样本,与理论值作比较,计算y=ax+b。(5)准确性:重复测定蒸馏水复溶后的FIB定值血浆、D-二聚体、FDP质控品(低值、高值)10次,计算均值和FIB的相对偏差。(6)正确度评价:选用国家卫生和计划生育委员会凝血5个批次室间质控品为正确度评价物,计算偏倚率。(7)生物参考区间的验证:选择20个健康体检人群进行生物参考区间转移的验证,计算R值。结果(1)批内不精密度:不同水平PT、APTT、FIB的CV分别为0.46%~1.06%、0.47%~2.66%、1.99%~3.92%,正常水平TT、D-二聚体、FDP的CV分别为0.71%、4.24%、7.05%。(2)日间不精密度:正常水平PT、APTT、FIB、TT的CV分别为2.62%、2.1%、8.9%、2.0%,异常水平PT、APTT的CV分别为2.85%、3.1%。(3)携带污染率:FIB为0.42%,D-二聚体为3.58%,FDP为1.96%。(4)检测线性范围:FIB为0.8~5.7 g/L,D-二聚体为0.50~30.00 mg/L,FDP为12.50~110.00 mg/L。(5)准确性:FIB相对偏差为7.15%,CV<15%,D-二聚体/FDP均在质控范围内。(6)生物参考区间的验证:总样本R值为0.96。(7)正确度评价结果:除一个批号的APTT偏倚处在上限外(15.28%),其余各指标均合格。结论本实验室新购的全自动血凝仪SYSMEX CA-7000各项性能指标基本符合要求,太阳试剂厂商提供的参考区间可以使用。 Objective To verify whether the performance indexes of the newly purchased SYSMEX automatic coagulation analyzer CA-7000 conform to the relevant standards and whether the reference interval provided by SUNBIO Shanghai can be directly applied to the clinic.Methods(1) Within-run imprecision:freshly mixed plasma [prothrombin time(PT),activated partial thromboplastin time(APTT),fibrinoge(FIB),three different concentration respectively;thrombin time(TT),D-Dimer,fibrin degradation product(FDP),normal concentration],measured 20 times in a row,then computed the standard deviation(SD),the coefficient of variation(CV).(2) Daytime imprecision:by using the quality control materials with two different concentration,measured once per day for 20 days,then calculate SD and CV.(3) Carryover rates:after the diluted plasma samples were measured by three times,the plasma samples in original concentration were measured three times again,then the first step was repeated,and then calculated carryover rate.(4) Linear range:measured the samples with five different concentrations and comparedthe measured value with theoretical values,then calculated y=ax+b.(5) Accuracy:dissolved by double distilled water,the standard plasma samples with fixed FIB value,D-Dimer and FDP quality control materials(low value and high value) were repeatedly measured 10 times,then calculate mean and relative deviation of the FIB.(6) Accuracy rating:choose 5 different batches of quality control material came from 5 different blood coagulation laboratories,which was accredited by NHFPC,as the correct assessment products to calculate bias.(7) Validation of biological reference interval:20 healthy volunteers were enrolled for validating the biological reference interval transfer,calculated R-values.Results(1) In-batch imprecision:in different level of PT,APTT and FIB,the CV were 0.46%-1.06%,47%-2.66%,and 1.99%-3.92%,respectively;in normal level of TT,D-Dimer and FDP,the CV were 0.71%,4.24%,and 7.05%,respectively.(2) Daytime imprecision:CV of normal level PT,APTT,FIB,and TT were 2.62%,2.1%,8.9%,and 2.0%,respectively;CV of abnormal level PT and APTT were 2.85% and 3.1%.(3) Carryover rates:FIB was 0.42%,D-Dimer was 3.58%,FDP was 1.96%.(4) Detection of linear range:FIB was 0.8-5.7 g/L,D-Dimer was 0.5-30.00 mg/L,FDP was 2.50-110.00 mg/L.(5) Accuracy:the relative deviation of FIB was 7.15%,CV〈 15%,the ratio of D-Dimer/FDP was in the context of quality control.(6) Validation of biological reference interval:the total sample R value was 0.96.(7) Accuracy evaluation results:all the indexes were qualified except that a bias from a batch of APTT which exceeded the upper limit.Conclusion The performance indicators of our new full-automatic blood coagulation analyzer,SYSMEX CA-7000,conform to the relevant standards;and the reference range provided by SUNBIO can be used directly.Key words
出处 《实用检验医师杂志》 2017年第4期220-224,共5页 Chinese Journal of Clinical Pathologist
关键词 血凝仪 太阳试剂 性能验证与评价 生物参考区间 Coagulation analyzer Reagent Performance verification Biological reference region
  • 相关文献

参考文献7

二级参考文献80

  • 1储靖,陈宁,杨程.钠钾镁钙葡萄糖注射液用于术前预扩容对凝血功能的影响[J].实用检验医师杂志,2011,3(3):186-187. 被引量:2
  • 2Fuentes-Arderiu X, Mas-Serra R, Alumg-Trulls A, et al. Guide- line for the production of multicentre physiological reference values using the same measurement system. A proposal of the Catalan Association for Clinical Laboratory Sciences [J]. Clin Chem Lab Med, 2004, 42(7):778-782.
  • 3马斌荣.医学统计学[M].5版.北京:人民卫生出版社,1999:22-25.
  • 4Paul SH, Amadeo JP. Reference intervals: a user' s guide [M]. AACC Press, 2005:31-36.
  • 5中华人民共和国卫生部.WS/T402-2012.临床实验室检验项目参考区间的制定[s].北京:中国标准出版社,2013.
  • 6中华人民共和国卫生部.中华人民共和国卫生行业标准.WS/T405-2012.
  • 7Starkey CR, Davies L, Hoyer JD, et al. Clinical manifestations of hemoglobin Chic0 at high altitude. J Pediatr Hematol Oncol, 2006,28 : 760-762.
  • 8张曼,主编.体检结果导读.北京化工出版社,2010,10-20.
  • 9Tamizhselvi R, Moore PK, Bhatia M. Hydrogen sulfide acts as a mediator of inflammation in acute pancreatitis: in vitro studies using isolated mouse pancreatic acinar cells [J]. J Cell Mol Med, 2007, 11 (2): 315-326. DOI: 10.111 l/j.1582-4934.2007.00024.x.
  • 10Sebag SC, Bastarache JA, Ware LB. Therapeutic modulation of coagulation and fibrinolysis in acute lung injmy and the acute respiratory distress syndrome [ J ]. Curr Pharra Biotechnol, 2011, 12(9) : 1481-1496.

共引文献66

同被引文献34

引证文献6

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部