注射用替莫唑胺零期临床试验中微量泵给药体外模型的设计 被引量:1

In vitro model of administration of temozolomide injection using micro-pumps during phase 0 clinical trial
下载PDF
导出
摘要 目的建立注射用替莫唑胺零期临床试验微量泵给药的体外模型,检测和校准微量泵,保证后续零期试验微剂量给药的准确性。方法在相同的实验环境下,对Ⅰ期临床实验研究室12台微量注射泵进行定量检测,采用注射用替莫唑胺零期试验相同的给药速度(0.8 ml/h)和给药时间(90 min),模拟零期试验的给药过程,对实际测量液体量与理论值进行对比和误差计算。所有试验重复3次,取3组数据平均值计算每台微量泵的误差值。结果 12台微量泵的误差值范围在±5%以内,根据国家计量规定和零期临床试验微剂量给药要求,表明本研究室的12台微量泵给药精确度高,可以用于后续零期试验给药。结论在整个实验的过程中,重视微量泵使用的相关因素,保证零期试验给药的准确性。 Objective To develop an in vitro model for temozolomide injection infused by micro-pumps and to assess the performance of micro-pumps so as to guarantee the accuracy of infusion during the phase 0 clinical trial. Methods Twelve micro-pumps were tested under the same conditions, involving temperature, humidity, operators and IV line of the same type. In order to simulate the infusion process of phase 0 clinical trial, the infusion speed and duration were set at 0.8 ml/h and 90 min respectively based on the phase 0 study protocol of temozolomide. Each pump was tested three times and the mean liquid amount was calculated to compare with the theoretical amount. Results The errors of the twelve miro-pumps, ranging from - 5%-5%, met the national measurement provision and requirements of micro-pumps used during phase 0 clinical trials, suggesting that the high accuracy of our micro-pumps could guarantee the subsequent administration in the clinical trials. Conclusion It is necessary to attach importance to the use of micro-pumps to ensure the accuracy of administration during the phase 0 clinical trials.
出处 《空军医学杂志》 2017年第6期395-397,共3页 Medical Journal of Air Force
关键词 零期临床试验 注射用替莫唑胺 微量泵 体外模型 phase 0 clinical trial temozolomide infused micro-pump ain vitro model
  • 相关文献

参考文献7

二级参考文献37

  • 1陈燕旸,肖笑.微量输液泵流速质控中探索的相关问题[J].中国医疗器械信息,2006,12(10):57-62. 被引量:7
  • 2Scheiner G, Sobel R J,Smith D E ,et al. Insulin pump therapy: guidelines for successful outcomesIJ]. Diabetes E duc, 2009,35 (Suppl2) : 29S-41S.
  • 3Amoore J,Dewar D,Ingram P,et al. Syringe pumps and start-up time: ensuring safe practice[J]. Nurs Stand, 2001,15 ( 17 ) :43-45.
  • 4Han PY, Coombes ID,Green B. Factors predictive of intravenous fluid administration errors in Australian surgical care wards[J]. Qual Saf Health Care,2005,14:179 - 184.
  • 5Heise D, Rathgeber J,Kettler D. Causes of failure and dangers in the use of motor driven infusion pumps. Accidental closure of the infusion system[J]. Anaesthesist, 1998,47 (1) : 54 - 58.
  • 6International Electrotechnical Commission. IEC60601 - 2 - 24: 1998. Medical electrical equipment Part 2- 24: Particular requirements for the safety of infusion pumps and controllers[S]. IEC Published, 1998 : 2.
  • 7中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB9706.27-2005.医用电气设备第2-24部分:输液泵和输液控制器安全专用要求[M].北京:中国标准出版社,2006:6.
  • 8储博慧.应用微量注射泵存在的问题与护理对策[J].医疗装备,2011,12(3):285-286.
  • 9李小敏,祝鸿雁,冯秦.微量注射泵持续滴注在人工气道湿化中的应用[J].临床研究,2012,3(11):158.
  • 10羽云燕,秦慧,李雪清,等.微量注射泵注射器原因报警的简易处理方法[J].中国保健营养(中旬刊),2012(5):396.

共引文献57

同被引文献8

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部