摘要
本研究先介绍金宝血透机温度系统的基本工作原理,后结合临床常见的故障代码,展开分析检修,并重点逆向画出继电器板电路原理图,为大家深入解决该类问题提供有效的经验和借鉴作用。
This paper first introduces the basic working principles of Gambro HM's TS, then finds out reasons for faults with error codes, and further reverse engineers the circuit diagram from relay board. It provides an effective experience and reference for us to solve these problems in depth.
出处
《医疗装备》
2018年第3期174-175,共2页
Medical Equipment
关键词
金宝
血透机
温度系统
检修
继电器板
Gambro
Hemodialysis machine(HM)
Temperature system(TS)
Maintenances
Relay board