硫酸沙丁胺醇吸入气雾剂递送剂量测定方法比较 被引量:6

Comparison of measurement methods of delivered-dose uniformity of salbutamol sulfate inhalation aerosol
原文传递
导出
摘要 目的:采用不同递送剂量(delivered-dose uniformity,DDU)测定装置,测定吸入气雾剂递送剂量均一性,比较测定结果差异,为选择吸入气雾剂递送剂量的测定方法提供参考。方法:采用DUSA管和改良单剂量玻璃取样装置测定硫酸沙丁胺醇吸入气雾剂的递送剂量均一性,分别测定3批样品各10支的标示首揿和末揿的递送剂量,比较不同装置的测定差异。结果:DUSA管测定结果均值为82.7~83.6μg,改良单剂量玻璃取样装置测定结果均值为92.4~94.6μg。结论:DUSA管测定递送剂量重现性良好,建议选择适当的装置测定递送剂量。取样揿数考察了气雾剂的瓶内和瓶间差异具有代表性。 Objective: To determine the difference in delivered-dose uniformity( DDU) of salbutamol sulfate inhalation aerosol measured by different devices,thus to provide basis for selection of method for DDU determination of aerosol inhalation. Methods: The dose at the beginning of use and the end of use were determined for 10 individual inhalers by DUSA and modified glass sampling device,and the results were compared. Results:The delivered-dose was 82. 7 ~ 83. 6 μg as determined by DUSA method,and 92. 4 ~ 94. 6 μg by the method using modified glass sampling device. Conclusion: The results of DUSA were more reproducible and less variable than the method using modified glass sampling device. The DDU of inhalers represent the difference of delivered-dose during the use of the same bottle and between different bottles.
出处 《中国新药杂志》 CAS CSCD 北大核心 2018年第4期405-408,共4页 Chinese Journal of New Drugs
基金 国家"重大新药创制"科技重大专项资助项目(2015ZX093001002009)
关键词 硫酸沙丁胺醇吸入气雾剂 递送剂量 DUSA管 改良单剂量玻璃取样装置 salbutamol sulfate inhalation aerosol delivered-dose uniformity DUSA tube modified glassampling device
  • 相关文献

参考文献3

二级参考文献21

  • 1陈桂良,侯惠民.以氯氟烃(CFC)和氢氟烷(HFA)为抛射剂的沙丁胺醇气雾剂的特性比较[J].药物分析杂志,2006,26(1):144-147. 被引量:9
  • 2牛秀华,岳志华,张启明.对现版药典几种吸入型气雾剂的质量评价及标准中存在问题的探讨[J].中国药事,2007,21(9):768-770. 被引量:2
  • 3Brambilla G, Church T, Lewis D, et al. Plume temperature emitted from metered dose inhalers [J]. Int J Pharm, 2011, 405 (1-2) : 9-15.
  • 4de Boer AH, Gjaltema D, Hagedoom P, et al. Characterization of inhalation aerosols: a critical evaluation of cascade impactor analysis and laser diffraction technique [J]. Int J Pharm, 2002, 249 (1-2) : 219-231.
  • 5Singh M, Misra C, Sioutas C. Field evaluation of a personal cascade impactor sampler (PCIS) [J]. Atmos Environ, 2003, 37 (34) : 4781-4793.
  • 6Steckel H, Mailer BW. Metered-dose inhaler formulations with beclomethasone-17,21-dipropionate using the ozone friendly propellant R 134a [J]. Eur JPharm Biopharm, 1998, 46(1): 77-83.
  • 7Taki M, Marriott C, Zeng XM, et al. Aerodynamic deposition of combination dry powder inhaler formulations in vitro: a comparison of three impactors [J]. Int J Pharm, 2010, 388(1-2): 40-51.
  • 8Brambilla G, Church T, Lewis D, et al. Plume temperature emitted from metered dose inhalers [J]. Int J Pharm, 2011, 405(1-2): 9-15.
  • 9Global Initiative for Asthma. Global strategy for asthma management and prevention. Available from: http:// www. ginasthma. org. Date last updated: 2012. Date last accessed r August 2013.
  • 10Dockhorn RJ. Wagner DE. Burgess GL. etal. Proventil HFA provides protection from exercise-induced bronchoconstriction comparable to proventil and ventolin. Ann Allergy Asthma Immunol , 1997; 79(1): 85-88.

共引文献13

同被引文献55

引证文献6

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部