摘要 室内装修造成的苯系物污染严重影响居民的生命健康。本研究设计了一种新型螺旋管状膜固相萃取进样装置,并与便携式单光子电离-飞行时间质谱仪(SPI-TOF MS)联用,用于空气中痕量苯系物的现场快速检测。螺旋的结构设计可以有效地增加管状膜的长度,在14min内即可完成苯、甲苯和二甲苯的检测,灵敏度分别提高了32.1、71.8和129.9倍,线性相关系数(R^2)均大于0.990 0,定量限(LOQ)分别低于33.3、67.8和93.6μg/m^3,能够满足国家规定的室内空气苯、甲苯和二甲苯的排放限值。重复测试8次,三者的相对标准偏差(RSD)分别为3%、1.7%和2.4%。该便携式质谱仪体积仅为(29×40×31)cm^3,质量为27kg,有望用于室内空气低浓度苯系物的现场多点快速检测。 Benzene series emitted from indoor decorating materials seriously affect people's health.In this study,an innovative helical membrane solid-phase extractioninlet was designed and coupled to a portable single photon ionization-time of flight mass spectrometry(SPI-TOF MS)for in situ and rapid detection of trace benzene series in air.A hollow fiber membrane(HFM)was applied to adsorb the benzene series,and the membrane material was polydimethylsiloxane(PDMS).In the membrane extraction device,the length of the HFM was prominently increased with the helical structural design of the HFM.The installation length of the HFM was as long as 150 cm,but the volume of the membrane extraction device was only(3.14×484×92)mm^3.The problem with traditional membrane inlet that had a small membrane area was solved.The operating parameters of the membrane extraction device were optimized,which mainly targeted for the sampling flow rate,desorption temperature and enrichment time.As a result,the optimal sampling flow rate was found to be 600 mL/min,the optimal desorption temperature was 180 ℃ and the enrichment time was selected as 5 min.Under the optimal conditions,compared with the direct capillary sampling,the detection sensitivity of the benzene,toluene and xylene were enhanced 32.1,71.8 and 129.9 folds within14 min analytical period,respectively.The concentration range of 69.6-3 482.1,82.1-4 107.1 and 94.6-4 732.1μg/m3 of the benzene,toluene and xylene standard gas were tested for quantitative analysis.As a result,the linear correlation coefficient(R^2)of the three target compounds were all greater than 0.990 0.Based on the standard curves and the signal-to-noise(S/N)ratios,the limits of quantitation(LOQs,S/N=10)of the benzene,toluene and xylene were calculated as 33.3,67.8 and 93.6μg/m^3,respectively,which were much lower than the legal limit of BTX in room air.In addition,3 482.1,4 107.1 and 4 732.1μg/m^3 standard samples for benzene,toluene and xylene were tested by this system for 8 cycles while the thermal desorption processes of each cycle were tracked.The relative standard deviation(RSDs)of eight individual analytical periods for the BTX were 3%,1.7%and 2.4%,respectively,showed good repeatability.Finally,the whole system was used for analyzing laboratory air,and characteristic peaks of toluene,xylene,acetone,butanone and ethyl tert butyl ether(ETBE),et al were recognized in the mass spectrum.Based on the standard curves,the concentration of the toluene and xylene were calculated as 73.6 and 100.5μg/m^3.The size of the portable SPI-TOF MS was only(29×40×31)cm^3 with a weight of 27 kg,which was prospectively applied for in situ and multipoint rapid detection of trace benzene series in indoor air.
出处 《质谱学报》 EI CAS CSCD 北大核心 2018年第2期141-148,共8页 Journal of Chinese Mass Spectrometry Society
基金 国家自然科学基金(21375129) 863专项(2014AA06A507)资助
关键词 螺旋管状膜固相萃取进样 单光子电离 便携式飞行时间质谱仪 苯系物 现场检测 helical membrane solid-phase extraction inlet single photon ionization portable time of flight mass spectrometry benzene series field test
  • 相关文献

参考文献6

二级参考文献104

共引文献63

同被引文献45

引证文献4

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部