设计制作了基于环绕式脉搏血氧探头的睡眠呼吸暂停综合征(SAS)检测装置原型,与临床上商用指夹式血氧探头进行了外周血氧饱和度(SpO2)测量的实验对比,并通过模拟呼吸暂停对SAS进行了初步检测和结果分析。数据表明,环绕式探头SpO2测量误差均在±1%以内,为SAS自动检测模型提供可靠的数据来源。此装置原型为实现SAS的自动筛查提供了一种新方法。
In this research we designed a prototype of Sleep Apnea Syndrome(SAS)detection device based on the surrounded type pulse oximeter probe.This self-designed device was compared with clinical commercial finger clip type oxygen probe.Furthermore,SAS was preliminary detected and analyzed simulated apnea.The results show that the peripheral oxygen saturation(SpO2)measurement error of this self-designed device is less than ±1%,which meets the requirements of sleep monitoring,thus,can provide reliable data sources for SAS automatic detection model.This device provides a new choice fro SAS screening.
出处
《吉林大学学报(工学版)》
EI
CAS
CSCD
北大核心
2018年第2期640-644,共5页
Journal of Jilin University:Engineering and Technology Edition
基金
吉林省自然科学基金项目(41174095)
关键词
仪器仪表技术
睡眠呼吸暂停综合征
脉搏血氧监测
脉率变异性
环绕式血氧探头
technology of instrument and meter
sleep apnea syndrome(SAS)
pulse oxygen saturation monitoring
pulse rate variability