摘要 针对CTBT放射性核素台站惰性气体氙β -γ符合谱能量漂移问题,提出了一种基于质量控制(QC)谱的谱能量漂移校正方法,利用质量控制谱中的γ能谱和二维β -γ符合谱分别对γ探测器和β探测器的能量漂移进行校正,确定氙样品二维β -γ符合谱中四种放射性氙同位素(131mXe,133mXe,133Xe和135Xe)的符合能量区域,实现了放射性氙同位素活度浓度的准确定量分析。 Aiming at the problem of energy shift of β -γ coincidence spectra from CTBT noble gas stations,an energy shift correction method based on QC spectrum was developed and implemented. Both the γ spectrum and β -γ coincidence spectrum of QC source are utilized. X/γ-rays in the γspectrum are utilized to recalculate the γ detector energy function. The coincidence regions of Compton scatters in β -γ coincidence spectrum is then used to recalculate the ! detector energy function. The coincidence regions of radio xenon isotopes(^131 mXe,^133 mXe,^133 Xe and^135 Xe) in β -γ coincidence spectrum of xenon sample are obtained accurately and the activity concentrations of radio xenon isotopes can be calculated accurately.
作者 张新军 王世联 贾怀茂 李奇 樊元庆 赵允刚 王军 侯龙 ZHANG Xin-jun1,2, WANG Shi-lian2 ,JIA Huai-mao2, LI Qi2, FAN Yuan-qing2 , ZHAO Yun-gang2 WANG Jun2, HOU Long1(1. China Institute of Atomic Energy, Beijing 102413, China; 2. CTBT Beijing National Data Centre and Radionuclide Laboratory, Beijing 100085, Chin)
出处 《核电子学与探测技术》 北大核心 2017年第8期824-829,共6页 Nuclear Electronics & Detection Technology
基金 国家自然科学基金青年基金资助项目(11505082)
关键词 全面禁止核试验条约 惰性气体氙 β-γ符合 能量漂移校正 Comprehensive Nuclear-Test-Ban Treaty noble gas xenon β-γ coincidence energy shift correction