能量响应修正法现场校准固定式γ剂量仪

On-Site Calibration of Stationary Gamma Dosimeter Using Energy Response Correction Method
下载PDF
导出
摘要 为解决大范围布放式探测器校准的难题,以便携式X射线为校准源,采用能量响应修正法对固定式γ剂量仪现场校准方法进行研究。在实验室对待校准同类型剂量仪进行了高能区和低能区的能量响应测试,结合蒙特卡罗模拟获取了剂量仪器能量响应曲线;在待标仪器现场,采用替代法对剂量仪器进行校准,通过已得到的能量响应曲线对校准系数进行修正,以获得最终的校准系数。比对实验结果表明,现场校准系数与实验室校准系数误差在5%以内,该方法提高了固定式γ剂量仪测量准确性,对于辐射安全管理和辐射环境评价具有重要意义。 In order to solve the difficult problem of calibration for large range radiation detectors,an onsite calibration method for stationary gamma dosimeter was studied using X ray source. The energy response of both high energy and low energy regions for dosimeter which was the same type as the on-site instrument were tested in the laboratory,and the energy response curve of the dosimeter was obtained using Monte Carlo simulation. Then the substitution method was used for dosimeter calibration in the field,and the calibration coefficients were modified with the energy response curve. The results of comparison experiment have shown that the error between on-site calibration coefficient and laboratory calibration coefficient was less than 5%. This method can improve the measurement accuracy of stationary dosimeter,and it has great reference value for the radiation safety management and radiation environment evaluation.
作者 吕汶辉 过惠平 吕宁 赵括 田晨扬 王晓添 侯毅杰 LV Wen-hui, GUO Hui-ping, LV Ning, ZHAO Kuo, TIAN Chen-yang, WANG Xiao-tian, HOU Yi-jie(Rocket Force University of Engineer, Xi'an 710025, Chin)
机构地区 火箭军工程大学
出处 《核电子学与探测技术》 北大核心 2017年第8期841-846,共6页 Nuclear Electronics & Detection Technology
关键词 X射线源 现场校准 能量响应 γ剂量仪 X-ray source on-site calibration energy response Gamma radiation dosimeter
  • 相关文献

参考文献6

二级参考文献44

  • 1于水,王功鹏,关守任.低大气层宇宙射线剂量水平及机组人员受照剂量的研究[J].辐射防护,1997,17(2):103-108. 被引量:7
  • 2岳清宇,金花.低大气层中宇宙射线电离量的分布测量[J].辐射防护,1988,8(6):401-409.
  • 3高飞.Co-60、Cs-137光子参考辐射场中散射辐射问题的蒙卡方法研究[D].北京:中国原子能科学研究院,2009.
  • 4Health and Sa{ety Laboratory. High pressure ar- gon ionization chamber systems for the measure- ment of environmental radiation exposure rates [R]. New York: HASL260, 1972.
  • 5任天山,林莲卿,陈志鹏.宇宙射线电离量的测定和几种探测器对宇宙射线的响应[J].辐射防护,1987,7(3):185-190.
  • 6International Organization for Standardization. X and gamma reference radiations for calibrating dose- meters and dose rate meters and for determining their response as a function of photon energy, Part 1: Radiation characteristics and production method [S]. Geneva: ISO Publication Office, 1997.
  • 7全国核能标准化技术委员会.GB/T12162.1-2000用于校准剂量仪和剂量率仪及确定其能量响应的x和Y参考辐射,第一部分:辐射特性及产生方法[s].北京:中国标准出版社,2000.
  • 8de PLANQUE G, GESELL T F. Environmental measurements with thermoluminescence doseme- ters-trends and issues[J]. Radiation Protection Dosimetry, 1986, 17: 193-200.
  • 9MAIELLO M L G. 8'h international intercompar- ison of environmental dosemeters[J]. Radiation Protection Dosimetry, 1990, 32: 91-98.
  • 10MAIELLO M L G. Ninth international 'mini' intereomparison of environmental dosemeters: A further analysis of the data[J], Radiation Protec- tion Dosimetry, 1995, 58: 167-175.

共引文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部