摘要
为了给激光尘埃粒子计数器提供可靠的校准、标定依据,研究0.296、0.498、1.019、2.995μm这4种标准粒子在不同稀释液喷射流量和粒子数下的分布规律,分析粒子发生系统中粒子数随工作时间的变化关系,探讨粒子计数器校准中的2种测量方法。结果表明,随粒子粒径的增加,测量粒子数所占总粒子数分数逐渐下降;前3种标准粒子具有稳定分布,后一种分布不稳定,且测量粒子数所占总粒子数分数与稀释液的喷射流量和浓度无关;粒子发生系统的稳定时间约为10 min;不同流量计数器比较校准时,前3种标准粒子适用于双管同时测量的方法,后一种适用于单管分时测量的方法。
In order to provide the reliable calibration rules for laser dust particles counter, the standard particles distribution rule of 0.296、0.498、1.019 and 2.995 μm were studied under the conditions of different diluents jet flow and particles concentration.The particles concentration as a function of working time was analyzed by the particles generation system. Two calibration methods of particles counter were discussed. The results show that the percentage of measured particles descends with the increase of particles size. The distributions of the first three standard particles are stable compared to the latter. The percentage of measured particles are independent of diluents jet flow and particles concentration. The settling time of particles occurrence system is about 10 minutes. The first three standard particles are suitable for double tube simultaneous measurement and the latter is suitable for single tube time-sharing measurement in the comparison calibration of different flow particles counter.
作者
詹永波
张健伟
曾建雄
石先城
ZHAN Yongbo;ZHANG Jianwei;ZENG Jianxiong;SHI Xiancheng(Guangdong Huankai Microbial Sci. &Tech. Co. ,Ltd., Guangzhou 510663, China)
出处
《中国粉体技术》
CAS
CSCD
2018年第3期69-75,共7页
China Powder Science and Technology
基金
广东省科学院引进领军人才项目
编号:2016GDASRC-0207
关键词
标准粒子
分布规律
粒子发生系统
双管同时测量
单管分时测量
standard particles
distribution rule
particles generation system
double tube simultaneous measurement
single tube time-sharing measurement