钴源后装机源强度实验测量与蒙特卡罗模拟计算的比较研究 被引量:3

A comparative study on the actual measurement and Monte Carlo simulation about radiation intensity of cobalt source of afterloader device
下载PDF
导出
摘要 目的:针对钴源医用后装机源强度,即源空气比释动能强度进行剂量学研究。方法:分别对国内临床使用的国产GZP3型钴源后装机和德国Co0.A86型钴源后装机,采用井型电离室实验测量和基于蒙特卡罗模拟计算的方法,进行源空气比释动能强度的剂量学比较研究。结果:实验测量和蒙特卡罗模拟计算的源空气比释动能强度偏差在3.0%之内,两种型号的钴源蒙特卡罗模型得到验证。结论:基于蒙特卡罗模拟的单位源活度空气比释动能强度因子的源强度表示可应用于钴源后装机质量控制检测中,本研究建立的两种型号的钴源蒙特卡罗模型可以开展相关临床剂量学研究应用。 Objective: To carry out dosimetry study for the intensity of primal air kerma of the cobalt source of medical afterloader device. Methods: The experimental measurement of well-type ionization chamber and Monte Carlo simulation method were adopted to compare the dosimetry of primal air kerma between GZP3 of domestic afterloader device of cobalt source and Co0.a86 of Germany. Results: The deviation of intensity of primal air kerma between the experimental measurement and Monte Carlo simulation was within 3%, and the Monte Carlo models of the two types of cobalt source were verified. Conclusion: The primal intensity of intensity factor of air kerma per unit source activity based on Monte Carlo simulation can be applied in the test of quality control of cobalt source of afterloder device. The two type of Monte Carlo models of cobalt source which established in this study can be used to carry out correlatively clinical research of dosimetry.
作者 袁继龙 马桥 何志坚 薛娴 程金生 YUAN Ji-long,MA Qiao,HE Zhi-jian(Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China)
出处 《中国医学装备》 2018年第5期18-20,共3页 China Medical Equipment
基金 国际原子能机构2014-2015国际合作项目(CPR6006)"加强现代核医学 放射诊断和放射治疗实践的质量保证"
关键词 近距离治疗 钴源 空气比释动能强度 后装治疗机 Brachytherapy Cobalt source Air kerma intensity Medical afterloader device
  • 相关文献

参考文献3

二级参考文献20

  • 1Richter J, Baier K, Flentje M. Comparison of 60Co bait and 192 Ir idium sources in high dose rate afterloading brachytherapy [J]. Strahlenther Onkol, 2008, 184 ( 4 ) : 187-192. DOI: 10. 1007/ s00066-008-1684-y.
  • 2Papagiannis P, Angelopoulos A, Pantelis E, et ai. Monte Carlo dosimetry of 60Co HDR brachytherapy sources [J]. Med Phys, 2003,30(4) :712-721.DOI : 10.1118/1.1563662.
  • 3Vijande J, Granero D, Perez-Caiatayud J, et ai. Monte Carlo dosimetrie study of the Flexisource Co-60 high dose rate source [J]..J Contemp Brachytherapy,2012,4( 1 ) : 34-44.DOI: 10.5114/ jcb.2012. 27950.
  • 4Granero D, P6rez-Calatayud J, Ballester F. Technical note: dosimetric study of a new Co-60 source used in brachytherapy [ J]. Med Phys, 2007,34 ( 9 ) : 3485 -3488. DOI : 10. 111 8/1.2759602.
  • 5Ballester F, Granero D, Perez-Calatayud J, et al. Monte Carlo dosimetric study of the BEB1G Co-60 HDR source [J].Phys Med Biol, 2005,50 ( 21 ) : N309-N316. DOI: 10. 1088/0031-9155/50/ 21/N03.
  • 6Taylor REP, Rogers DWO. An EGSnrc Monte Carlo-calculated database of TG-43 parameters [ J ] .Med Phys ,2008,35 (9) :4228- 4241. DOI : 10.1118/1.2965360.
  • 7Selvam TP, Bhola S.Technical note : EGSnrc-based dosimetric study of the BEBIG 60Co HDPt brachytherapy sources [J].Med Phys, 2010,37 ( 3 ) : 1365-1370.DOI : 10.1118/1.3326948.
  • 8Perez-Calatayud J,Ballester F, Das RK,et al. Dose calculation for photon-emitting brachytherapy sources with average energy higher than 50 keV: report of the AAPM and ESTRO [J]. Med Phys, 2012,39 ( 5 ) :2904-2929. DOI : 10.1118/1.3703892.
  • 9Pujades-Claumarchirant MC, Granero D, Perez-Calatayud J, et al. Evaluation of interpolation methods for TG-43 dosimetric parameters based on comparison with Monte Carlo data for high- energy brachytherapy sources [J]. J Contemp Brachytherapy, 2010,2( 1 ) :28-32.DOI : 10.5114/jcb.2010. 13715.
  • 10Borg J, Rogers DWO. Spectra and air-kerma strength for encapsulated t921r sources [ J]. Med Phys, 1999,26 ( 11 ) : 2441- 2444.DOI : 10.1118/1.598763.

共引文献7

同被引文献21

引证文献3

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部