2016年宝鸡市部分医用X射线诊断设备质量控制检测结果 被引量:4

Quality control test results of some medical X-ray diagnostic equipments in Baoji City in 2016
原文传递
导出
摘要 目的了解宝鸡市医用X射线诊断设备应用质量现状。方法使用B-Piranha型X射线性能检测仪及其配套设备,按照国家标准要求,检测透视机、DR、CR、CT机的性能指标。结果 2016年宝鸡市60台透视机、DR、CR、CT机性能指标整机初检合格率分别为85.71%、45.83%、23.08%、88.89%。透视机透视受检者入射体表空气比释动能率典型值、高对比分辨力、低对比分辨力合格率最高,达到100%;影像增强系统亮度自动控制合格率较低,为80.00%。DR机有用线束垂直度偏离的合格率最高,达到100%;SID偏离合格率最低,仅有45.83%。CR机输出量重复性、线性、有用线束垂直度偏离、曝光时间偏离合格率最高,均达到100%;SID偏离、IP响应均匀性及一致性合格率最低,为45.83%和23.08%。CT机诊断床定位精度、定位光精度、CT值、均匀性、线性合格率均达到100%,高对比分辨力最低,合格率仅有88.89%。结论宝鸡市医用X射线诊断设备初检合格率低,应加强放射诊疗设备的应用质量状态监测和监督。 [Objective]To know the quality situation of medical X-ray diagnostic equipments in Baoji City. [Methods] According to the national standard,B-Piranha X-ray tester and its corollary equipment were used to measure the fluoroscopic machine,DR(Digital Radiography) machine,CR(Computed Radiography) machine and CT(Computed Tormography) machine. [Results] The preliminary qualified rate of 60 fluoroscopic machines,DR machines,CR machines and CT machines in Baoji City was 85.71%,45.83%,23.08% and 88.89%,separately. The qualified rates of typical number of entrance surface air kerma on clients,highcontrast definition and low-contrast definition of fluoroscopic machine were the highest,which reached 100%,while the qualified rate of automatic brightness control in image enhanced system was lower,which was 80.00%. The qualified rate of vertical deviation of useful wiring harness of DR machine was the highest which reached 100%,while the qualified rate of SID deviation was the lowest,which was 45.83%. As for CR machine,the qualified rate of replicability,linearity,vertical deviation of useful wiring harness and deviation of exposure time were the highest,which reached 100%,while the qualified rates of SID deviation,and IP response homogeneity and conformity were the lowest,which was 45.83% and 23.08% separately. The qualified rates of positioning accuracy of diagnostic bed,localization light accuracy,CT value,heterogeneity and linearity all reached 100%,while the qualified rates of high-contrast definition was the lowest,which was 88.89%. [Conclusion] The preliminary qualified rate of medical X-ray diagnostic equipments is low in Baoji City. It is necessary to strengthen the monitoring and supervision on the application quality situation of radiation diagnostic equipments.
作者 刘红英 路建超 杨海峰 康志东 晁晖 槐玺龙 郭强 LIU Hong-ying;LU Jian-chao;YANG Hai-feng;KANG Zhi-dong;CHA;HUAI Xi-long;GUO Qiang(Labor and Radiation Health Department of Baoji City Center for Disease Control and Prevention,Baoji Shoanxi,721016,China)
出处 《职业与健康》 CAS 2018年第8期1017-1019,1023,共4页 Occupation and Health
关键词 X射线机 性能检测 质量控制 分析 X-ray equipment Performance detection Quality control Analysis
  • 相关文献

参考文献13

二级参考文献96

共引文献58

同被引文献21

引证文献4

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部