导出
摘要 直接观测太阳的星上定标方案中,设计了直径分别为20 mm和0.5 mm的光阑用于切换观测地球和太阳,两个光阑面积比的高精度测量是保证定标精度的关键。论述了光阑面积的测量原理,基于光阑面积测量的特点,利用激光点阵扫描法和通量比较法分别测得两个光阑与参考光阑的面积比,最终得出Φ20mm和Φ0.5 mm光阑的面积比值为1604.381,评估合成不确定度为0.046%.为光阑大面积比的高精度测量提供了可行方案,支持直接观测太阳的星上定标方法。 In the project of on-orbit calibration, two apertures were designed with nominal diameters of 20 mm and 0.5 mm for switching observations of the earth and the scen, the high precision measurement of the two apertures area ratio is the key to ensure the calibration accuracy. The measurement method of the area of aperture was discussed briefly. Based on the measurements characteristics of the area of aperture, the ratio of the area of the two apertures to the area of the referenced aperture was measured by utilizing laser raster scanning and flux comparison. The area ratio of aperture Φ20 mm to Φ0.5 mm was 1604.381 with a total relative combined standard uncertainty 0.046%. The conclusion provides a feasible solution for the high-precision measurement of the aperture area ratio, and supports the on-orbit calibration of direct solar viewing.
作者 赵文贤 李新 张艳娜 ZHAO Wenxian;LI Xin;ZHANG Yanna(Key Laboratory of Optical Calibration a~d Characterization, Anhui Institute of Optics and Fine Mechanics Chinese Academy of Sciences, Hefei 230031, China;University of Science and Technology of China, Hefei 230031, China)
出处 《大气与环境光学学报》 CAS CSCD 2018年第3期226-232,共7页 Journal of Atmospheric and Environmental Optics
基金 国家自然科学基金 11204318 61275173 41271373 安徽省自然科学基金 1508085QD80~~
关键词 直接观测太阳 点阵扫描 通量比较 面积比值 direct observations of the sun raster scanning flux comparison area ratio
  • 相关文献

参考文献4

二级参考文献27

  • 1盛建军,张黎明,李双,李健军,赵强,邱成风,田光.光阑面积的精确测量方法[J].大气与环境光学学报,2006,1(5):117-120. 被引量:1
  • 2李双,王骥,章骏平,吴浩宇,郑小兵.可见-近红外(488~944nm)基于低温辐射计的高精度光辐射绝对定标研究[J].光学学报,2005,25(5):609-613. 被引量:17
  • 3邾继贵,王大为,王鑫,叶声华.光学坐标测量系统中的关键技术[J].红外与激光工程,2007,36(3):296-299. 被引量:19
  • 4Harder J W, Fontenla J, Lawerance G, et al. The spectral irradiance monitor: measurement equation and calibration [J]. SoL Phys., 2005, 230(12): 169-204.
  • 5Fowler J B, Dezsi G. High accuracy measurement of aperture area relative to a standard known aperture [J]. Journal of Research of the National Institute of Standards and Technology, 1995, 100(3): 277-283.
  • 6JohnsonB C, Litorja M, Butler J J. Preliminary results of aperture area comparison for exoatmospheric solar irradiance [C]. Proc. SPIE, 5151: 455-462.
  • 7Hartmann J. Advanced comparator method for measuring ultra-small aperture areas [J]. Meas.Sci. Tech., 2001, 12(10): 1678-1682.
  • 8Fischer J, Stock M. A non-contact measurement of radiometric apertures with an optical microtopography sensor [J]. Meas. Sci. Tech, 1992, 3(8): 693-698.
  • 9Li Zhengqiang, Blarel Lue, Podvin Thierry, et al. Applied Optics, 2008, 47(10): 1368.
  • 10Beat Schmid, Christoph Wehrli. Applied Optics, 1995, 34(21): 4500.