摘要
对一种地表水中锑的快速检测技术进行实际应用。结果表明,该技术的线性关系良好,检出限为0.50μg/L,准确度和精密度良好,加标回收率为86.0%~120%。对干扰元素的影响进行分析,消除了20倍铜的干扰。与电感耦合等离子体质谱法的进行比对,结果没有显著性差异。该技术的操作简便、成本较低、结果可靠,既可在实验室内快速分析检测,也可在地表水污染事故现场进行快速检测。
The results showed the linearity relationship were excellent,the detection limit was 0.50μg/L,the accuracy and precision was good,the recoveries obtained by standard addition ranged from 86.0% to 120%.The influence of the interfering elements was analyzed,and the interference of 20 times copper was eliminated.Compared with inductively coupled plasma mass spectrometry,the results of statistics t test showed no significant difference between them.This technology is simple,fast,cheap,reliable,can be used in the laboratory or in the scene.
作者
林君铿
刘胜玉
梁永津
党永锋
Lin Junkeng;Liu Shengyu;Liang Yongjin;Dang Yongfeng(Monitoring Centre of Pearl River Valley Aquatic Environment, Cruangzhou 510611 ,China)
出处
《分析仪器》
CAS
2018年第3期124-127,共4页
Analytical Instrumentation
关键词
地表水
锑
快速检测
Surface water
Antimony
Rapid detection